4�6�3 Dc Powered Analyzer; Wiring - Input/Outputs; Figure 13 - Ac Terminal Board - Located In Rear (Wiring) Compartment; Figure 14 - Dc Terminal Board - Located In Rear (Wiring) Compartment - Ametek OXYvisor Installation, Operation & Maintenance Manual

Optical oxygen analyzer for use with bosx optical sensors
Table of Contents

Advertisement

Section 4
Installation
4�6�3 DC Powered Analyzer
Analyzer - Power Supply Connections (DC Version)
The OXYvisor control unit can operate using 24 VDC +/-10%.
must be from a Class2 or LPS source, OR that a fuse (1 A) will be required in the installation.
Please note that the 24VDC version of the analyzer is a 4-wire device. It has
separate wiring for power and signal (not a loop powered device).
USEFUL INFORMATION
DC Power
OXYvisor analyzers ordered with DC power input will have a termination board labeled with "24VDC (+)",
"24VDC (-)", and "Earth" (Ground) as shown in Figure (10).
4�7 Wiring - Input/Outputs
Input / Output Wiring
All Input / Output wiring for the OXYvisor analyzer is located on the left side of the board as shown in Figure (9)
and (10). The wiring assignments are shown in Table (1).
Input /Output wiring requirements:
!
Terminal
AC OPTION

Figure 13 - AC Terminal Board - Located in Rear (Wiring) Compartment

AC TERMINAL BOARD POWER WIRING
Label
Terminal
Line
Power 120/240VAC
85-264 VAC 47/63Hz Input Power
Neutral
Power 120/240VAC
85-264 VAC 47/63Hz Input Power
Earth
Power 120/240VAC
Table 6- AC Terminal Board Power Wiring
30
Installation, Operation and Maintenance Manual
26 to 16 gauge (IEC .400 to 1.25)
Copper, stranded wire
Minimum 300V Insulation
Tightened (torque) to .22 to .25 Nm
Terminal
Description
Function
Line Power wiring
Neutral Power Wiring
Earth Terminal
Earth Terminal
OXYvisor Optical Oxygen Analyzer
When energized by mains the 24 V DC supply
Terminal
DC OPTION

Figure 14 - DC Terminal Board - Located in Rear (Wiring) Compartment

DC TERMINAL BOARD POWER WIRING
Label
Terminal
Description
24VDC (+) Power 24VDC
21.6 - 26.4 VDC
24VDC (-) Power 24VDC
21.6 - 26.4 VDC
Earth
Power 24VDC
Earth Terminal
Table 7- DC Terminal Board Power Wiring
Terminal
Function
Positive DC Power wiring
Negative DC Power Wiring
Earth Terminal