Ametek OXYvisor Installation, Operation & Maintenance Manual page 66

Optical oxygen analyzer for use with bosx optical sensors
Table of Contents

Advertisement

Help Error
Codes
Error
X-type (critical) Error Description
Code
X21
Signal intensity is too high.
X22
Signal is too high.
X23
Spectrometer reference path error.
X24
Spectrometer EEPROM has failed.
X31
Spectrometer microprocessor has failed.
X32
Spectrometer pulse counter overflowed.
The error indicates that the firmware present in the spectrometer
X33
has undergone a reset.
This error indicates that some component on the main board has
failed. The error is never cleared if set. A number is printed on the
main display next to the X34 error. This number can be used to
identify the cause of failure. The bitmap shown below can be used
to identify the component that failed.
Bit 0 - ADC Communication Failure
Bit 1 - Altimeter wrong manufacture ID
Bit 2 - Altimeter Disconnected
Bit 3 - DAC 1 Communication Failure
Bit 4 - DAC 2 Communication Failure
Bit 5 - Reserved
X34
Bit 6 - Non Volatile Memory ID Failure
Bit 7 - Non Volatile Memory Communication Failure
Bit 8 - Flash ID Failure
Bit 9 - Flash Memory Type Failure
Bit 10 - Flash Memory Capacity Failure
Bit 11 - Flash Communication Failure
Bit 12 - RTD ADC Communication Failure
Bit 13 - UI Board Reset Error
Bit 14 - Flash Data Corruption Error
The error indicates that the system has encountered a power failure.
X55
This error is cleared once the user acknowledges this . This error is
not set for software resets.
Table 15.2 - "X" Critical - Error Code Table (Part 2)
66
Installation, Operation and Maintenance Manual
OXYvisor Optical Oxygen Analyzer
Trouble shooting and things to check
Signal intensity is defaulted for FlexSens style sensors. If you have a Fibersense, turn down the signal intensity. See Table 10
for general recommended values.
Record all errors and consult factory.
Power OFF and ON. If error is still present, consult the factory.
Record the details of error, "X34 - Bit YZ" and consult the factory
Indicated the analyzer lost power. Check wiring. Check power source.