Type 42 probe tube and filter unit installation

Before the installation
Type 42 probe tube
Installation
GAA330-M EMISSION MONITORING SYSTEM
CAUTION
The weight of the probe tube with filter unit amounts to approx.
28 to 32 kg! Two persons are needed for transportation and mounting!
Observe the "Piping plan" in the drawings set.
Make sure that the wall tube is installed at the extraction point.
(dimensions in mm)
1
Insert the probe tube in the wall tube and screw the probe tube flange
to the wall tube flange. Use the green seal from the accessories pack to
seal the space between the flanges.
2
Screw the filter unit to the flange of the probe tube. Use the green seal
from the accessories pack to seal the space between the flanges of
probe tube and filter unit.
3
Mount the heating sleeve on the filter unit.
4
If applicable, install the compressed-air hoses between the filter unit
and the back-purging unit.
|
OI/GAA330-M-EN REV. C