8
Safety labels affixed to the analyzer system
National regulations
Device safety and safe operation
Working with hazardous gases
Protective lead connection
GAA330-M EMISSION MONITORING SYSTEM
Observe the safety labels affixed to the analyzer system or to the individual
components:
Consult documentation!
Hot surface! (Temperature > 60 °C)
Corrosive material!
Risk of electric shock!
The regulations, standards and guidelines cited in this operating instruction
are applicable in the Federal Republic of Germany. The applicable national
regulations should be followed when the device is used in other countries.
The device is designed and tested in accordance with safety standard
EN 61010 Part 1 "Safety requirements for electrical equipment for measure-
ment, control, and laboratory use" and is shipped ready for safe operation.
To maintain this condition and to assure safe operation, read and follow the
safety information in this operating instruction and use the described ac-
cessories as specified in this operating instruction. Failure to do so can put
persons at risk and can lead to device damage as well as damage to other
systems and devices.
Some gas components whose concentration is measured with the analyzer
system are hazardous to health. For this reason, the sample gas must under
no circumstances be allowed to escape from the sample gas path in either
the measurement mode or when performing maintenance. The analyzer
system must be checked for leaks regularly. The diluted stack gas must be
discharged outside the room where the analyzer cabinet is installed. Ensure
adequate ventilation of the room in which the analyzer system is installed.
The legal requirements for the maximum work place limit values of the
measurement and test gases must be observed.
The protective lead (ground) should be attached to the protective lead con-
nector before any other connection is made.
|