18
Ambient temperature
Relative humidity
Dimensions and space requirement
Installation site stability
Sample gas inlet conditions (at the extraction point)
Power supply requirements
Service socket
GAA330-M EMISSION MONITORING SYSTEM
Operation
Storage and transport
Year-round average max. 75%, short-term max. 95%, occasional slight con-
densation is permitted.
Refer to the "Layout Plan" in the drawings set.
The installation site floor must be plane and the wall capable of supporting
the cabinets weight.
CAUTION
The analyzer system must not be used to measure flammable gases or
combustible gas/air or gas/oxygen mixtures!
Temperature
Max. 500 °C
Pressure p
850 to 1100 hPa (0.85 to 1.1 bar)
abs
Flow
Max. 200 l/h
Operating voltage
Power consumption
Provide a mains isolator or a switched socket-outlet in order to be able to
disconnect all the power from the analyzer system if required.
230 VAC, 48 to 62 Hz, max. 5 A.
The service socket is located on the right side of the cabinet; access is given
from the outside.
|
OI/GAA330-M-EN REV. C
with ventilation fan
with cooling unit
after draining and drying parts in
contact with condensate
230 VAC ± 10%, 50 to 60 Hz ± 3 Hz;
L1, L2, L3 each max. 25 A
Basic version (incl. AO2020-Uras26)
Cooling unit
Filter unit type PFE3 or PFE2, heated
Probe tube type 42, heated
Sample gas line type TBL01-S
+5 to +35 °C
+5 to +45 °C
+2 to +60 °C
−25 to +60 °C
950 VA
+ 410 VA
+ 250 VA
+ 800 VA
+ 90 VA/m