66
Visual inspection
Cleaning hints
GAA330-M EMISSION MONITORING SYSTEM
Device, module
Gas analyzer display: Measured values,
1
status messages
Condensate bottle status indicator
2
Cabinet temperature indicator
3
4
Bypass flow meter
Analyzer flow meter
5
Heated probe temperature controller
6
Measuring point and back-purging
7
pushbuttons
Main switch
8
Heated line temperature controller
9
Status indicators: Temperature alarm,
10
Fault cooler, Fault humidity, Flow
alarm min, Flow alarm max
Sample gas cooler temperature
11
Cooling unit
Never use water or any solvents to clean parts inside the analyzer cabi-
net.
Always operate the analyzer system with cabinet door closed. Remove
dust inside the analyzer cabinet using a broom and a vacuum cleaner.
Clean the outside of the analyzer cabinet with a wet towel and mild
cleaning agents. Pay attention that no droplets invade the cabinet.
|
OI/GAA330-M-EN REV. C
Nominal condition
Green LED "Power" ON
LED OFF
< 45 °C
120 l/h
40 to 80 l/h
180 °C
Measuring point pushbutton
ON, Back-purging pushbutton
OFF (can be pushed during
visual inspection)
ON
180 °C
All LEDs OFF
5 °C
Typical set point: 35 °C