Electrical Connection; Safety Information; Notes - ABB GAA330-M Operating Instructions Manual

Emission monitoring system for marine applications
Table of Contents

Advertisement

36

Electrical connection

Safety information

Notes

GAA330-M EMISSION MONITORING SYSTEM
CAUTION!
Please observe the relevant national safety regulations for the construc-
tion and operation of electrical installations as well as the following safe-
ty instructions.
Before connecting the power supply, check that the operating voltage on
the rating plate is the same as the mains voltage.
The protective-conductor terminal must be connected to a protective
conductor before any other connections are set up.
The device can be dangerous if the protective conductor is interrupted
inside or outside the device or the protective-conductor terminal is dis-
connected.
CAUTION!
An easily accessible mains isolator must be installed near the analyzer
cabinet in order to be able to completely disconnect the latter from the
power supply. Mark the mains isolator in such a way that assignment to
the device to be separated can be clearly identified.
NOTE
Lay the signal lines separate from the power supply lines. Carefully plan the
combination of signal lines in cables.
NOTE
The analyzer system must be hardwired!
NOTE
When installing the electrical connections, comply with the information con-
tained in the drawings set. The connection drawings in this chapter are only
examples.
|
OI/GAA330-M-EN REV. C

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents