Program sequence

Program sequence
PFE2
Digital
output:
Valve:
Step Duration Function
0
8 h
1)
Measure
Back-purging
1
10 s
6)
probe filter
Back-purging
2
14 s
7)
probe tube
3
6 s
Venting
4
150 s
2)
Post-purging closed
0
8 h
1)
Measure
PFE3
Digital
output:
Valve:
Step Duration Function
0
8 h
1)
Measure
1
2 s
Switch over
Back-purging
2
4 s
3)
probe filter
Back-purging
3
8 s
4)
probe tube
4
6 s
Venting
5
150 s
2)
Post-purging
0
8 h
1)
Measure
GAA330-M EMISSION MONITORING SYSTEM
-D08
-D08
DO2
DO3
-Y1.1
-Y2.1
Impulse
Filter
compr.
back-
air
purging
closed
closed
Impulse open
Impulse closed
closed
open
closed
closed
closed
-A01
DO1
-Y12
Filter
back-
purging
closed
closed
open
closed
closed
closed
closed
1) Cycle time factory-set to 8 hours
2) Post-purging time factory-set to 150 s
3) 1x pressure impulse 2 s, 1x interrupt 2 s
4) 1x pressure impulse 2 s, 1x interrupt 2 s, 1x pressure impulse 4 s
5) Calibration valve on = "Measure", calibration valve off = "Calibrate"
6) 3x interrupt 2 s, 2x pressure impulse 2 s
7) 3x interrupt 2 s, 2x pressure impulse 2 s, 1x pressure impulse 4 s
|
OI/GAA330-M-EN REV. C
-D08
-D08
-E05
DO1
DO4
MV1
-Y3.1
-Y4
-Y01
Tube
Position
back-
calibr.
purging Venting
valve
closed
open
Measure off
closed
closed
Calibrate on
open
closed
Calibrate on
closed
open
Calibrate on
closed
open
Measure on
closed
open
Measure off
-A01
-A01
-E05
DO2
DO3
MV1
-Y11
-Y07
-Y01
Tube
Position
back-
calibr.
purging Venting
valve
closed
closed
Measure off
closed
closed
Calibrate on
closed
closed
Calibrate on
open
closed
Calibrate on
closed
open
Calibrate on
closed
closed
Measure on
closed
closed
Measure off
47
Status
Display
signal
Message
"Purge
back
Maint.
5)
active"
mode
off
on
on
on
on
off
Status
Display
signal
Message
"Purge
back
Maint.
5)
active"
mode
off
on
on
on
on
on
off