Replacing the cooling unit / fan outlet filter mesh

When is filter mesh replacement needed?
Replacing the filter mesh

Replacing the suction filter in the diaphragm pump

When does the suction filter need to be replaced?

Emptying the condensate collection bottle

Emptying the condensate collection bottle
GAA330-M EMISSION MONITORING SYSTEM
The cooling capacity of the cooling unit or the ventilation fan depends upon
the cleanness of the filter mesh. It should be replaced if it begins to turn
dark.
1
Remove the grid which holds the filter mesh in place.
2
Change the filter mesh.
3
Re-assemble the grid.
CAUTION
Residues of the gas that the pump has been feeding may be found on the
diaphragm and valve plates. There materials can flow out when the dia-
phragm pump is opened. Take appropriate measures where needed to
collect such residues.
The suction filter in the pump must be replaced when the pump is no longer
feeding gas efficiently enough.
CAUTION
When working with corrosive reagents note the hazard information and
safety precautions contained in the applicable material safety data
sheets.
Condensate is often acidic. Neutralize condensate and follow the pre-
scribed measures for disposal.
Empty the condensate collection bottle when the applicable status message
is displayed.
|