Troubleshooting
System Recovery
HDD Revision: NM320 or below
You need a keyboard for this operation.
Step 1. Shut down the E5052B.
Step 2. Disconnect all of the USB device from the USB ports.
Step 3. Connect the keyboard to the E5052B.
Step 4. Press the standby switch of the E5052B to turn it on.
Step 5. When the screen as shown in the figure below appears, press
keyboard.
e5071cse1055
Press F11 only once. If you press F11 twice or more and a screen which is
not mentioned in this manual appears, reboot the instrument and restart
the process from the first.
After several seconds, the next screen appears automatically even if you
do not press any key, so do not miss it.
After
power off and remove unnecessary USB devices. Then, replace the
keyboard with a PS/2 keyboard to execute the process again.
Keysight E5052B Signal Source Analyzer
Starting Acronis Loader...
Press F11 for Acronis Startup Recovery Manager...
Step
5, if the E5052B freezes with the cursor does not blink, turn the
of the
F11