Troubleshooting
System Recovery
Step 9. After the restart, execute initial registration. For information on the execution
procedure, refer to
Never turn off the power during the system recovery because doing so
may cause serious damage to the E5052B.
Procedure for Windows Vista Business
HDD Revision: NM32x
Only the recovery function for Factory image is loaded.
The User backup function is not available.
You need a keyboard for this operation.
Step 1. Turn ON the E5052B.
Step 2. The boot option is displayed. Use the downward arrow key to select the
"Keysight Recovery Solution" and press Enter
Step 3. Windows PE starts up and "Keysight Recovery Program" window is displayed.
Press 1, and then press Enter.
Figure 3-6
Keysight Recovery Program(1/1)
Step 4. A warning message is displayed. Press 1, and then press Enter.
42
"Initial Registration of E5052B" on page
26.