Installation
Installing Front Handles/Rack Mounting Flanges

How to Install the Rack-mount Kit (Option 1CM)

The rack-mount kit includes two flanges (locking side plates) for mounting the
E5052B on a rack (482.6 mm/19 inches) conforming to the EIA Standard.
While referring to
steps.
Step 1. Remove the adhesive-backed trim strip (1) from each side of the outer frame
of the E5052B front panel.
Step 2. Use the provided screws to mount a rack-mounting flange (4) on each side of
the E5052B front panel frame.
Step 3. Remove the four bottom feet of the E5052B (lift the bar marked TAB on the
inner side of the foot and slide the foot toward the bar).
Step 4. Mount the E5052B on the rack.

How to Install the Rack-mount and Handle Kit (Option 1CP)

The rack-mount and handle kit includes both the rack-mounting flanges
(locking side plates) and front handles. While referring to
rack-mount kit by following these steps.
Step 1. Remove the adhesive-backed trim strip (1) from each side of the outer frame
of the E5052B front panel.
Step 2. Use the provided screws to mount a front handle (2) and rack-mounting
flange (4) on each side of the E5052B front panel frame.
Use both a front handle and a rack-mounting flange in the same time
certainly. Do not attempt to install flanges or handles separately with
hardware provided. Serious electrical damage will occur to the instrument.
Step 3. Remove the four bottom feet of the E5052B (lift the bar marked TAB on the
inner side of the foot and slide the foot toward the bar).
Step 4. Mount the E5052B on the rack.
18