■ Setting the median filter
The intermediate value of multiple sets of data can be output as the measurement
result.
Setting [Display]
Low [LOW]
Mid [MID]
High [HIGH]
Off [OFF]
(Default value)
As an example, here is an explanation of the procedure for selecting the "Mid" median
filter setting.
Operating procedure
Steps
Key operation
1
RUN
RUN
FUN
FUN
L
TEACH
TEACH
2
ZERORST/ ESC
ZERORST/
3
ZERO/
ZERO/ SET
4
5
6
ZERORST/
ZERORST/ ESC
ZERO/
ZERO/ SET
7
RUN
RUN
FUN
FUN
L
TEACH
TEACH
Outputs the intermediate value of the last 3 measurements.
Outputs the intermediate value of the last 9 measurements.
Outputs the intermediate value of the last 15 measurements.
Median filter is not used.
Display
UNLIT
H
L
RUN
RUN
ESC
SET
ESC
SET
LIT
H
L
RUN
RUN
Setting the Output Conditions
Description
Description
RUN
RUN
Press
key for two seconds to enter the FUN
FUN
FUN
L
TEACH
TEACH
mode.
Press
or
keys to select either of "MEAS" and
press
key.
ZERO/ SET
ZERO/
SET
Press
or
keys to select either of "TASK1"
and press
key.
ZERO/ SET
ZERO/
SET
Press
or
keys to select either of "FILTER"
and press
key.
ZERO/ SET
ZERO/
SET
Select "Median filter" as the filter type.
Press
or
keys to select either of "MEDIAN"
and press
key.
ZERO/
ZERO/ SET
SET
Select "Mid" as the median filter type.
The current setting value is displayed on the sub-
display.
Press
/
keys to enter the editing mode, and the
sub-display blinks.
Select "MID" and press
ZERO/ SET
ZERO/
SET
RUN
RUN
Press
key for two seconds to enter the RUN
FUN
FUN
L
TEACH
TEACH
mode.
Chapter 3
key.
ZW
User's Manual
3-15