● Measured value 2 mode (When measured values for multiple tasks are output by
BINARY_SEL input)
Judgment output
(HIGH/PASS/LOW)
TASK A
measurement value
TASK B
measurement value
BINARY_SEL input
Binary output
ON
GATE output
OFF
BINARY OUT output
T0
Measurement cycle
T4
GATE signal rise time
T5
Output time of GATE signal
T15
Binary output update cycle
T17
Response time of input
• Explanation of operations (When the results for all tasks are output)
(1)
Input the number of the task you want to output to the BINARY_SEL input.
(2)
The results for the selected tasks are output (judgment output and binary
output).
(3)
After the judgment result and binary data output, the GATE signal is turned ON.
(4)
The GATE output time after the GATE signal is turned ON, the GATE signal is
turned OFF.
(5)
When the results for the selected tasks are updated, the judgment output and
binary output are also updated.
(6)
Input the number of the task you want to output to the BINARY_SEL input.
(7)
The results for the selected tasks are output (judgment output and binary
output).
(8)
After the judgment result and binary data output, the GATE signal is turned ON.
(9)
The GATE output time after the GATE signal is turned ON, the GATE signal is
turned OFF.
(10) When the results for the selected tasks are updated, the judgment output and
binary output are also updated.
T15
T4
T17
T0
T5
TASK A
Item
0.5 ms
-
Setting value
(0.1 to 100.0 ms)
T0uUpdate cycle
-
T17 T4
T5
TASK A result
TASK A result
TASK A
Minimum
Depends on the set
conditions
0.2 ms
Setting value
(0.1 to 100.0 ms)
T0uUpdate cycle
T0
Chapter 4
Timing Charts
TASK B
TASK B result
TASK B result
TASK B
Maximum