● Two-point scaling
A measurement is performed at two positions and offset values are set for that
measured value. (Default value: OFF)
Example: When correcting display values to match actual distances
Separate the two specified points by at least 1 % of the rated measuring range for the connected
Sensor Head.
For example, for ZW-S40, the two measured points must be separated by at least 12 mm x 0.01 =
0.12 mm as the measuring range is 12 mm u 6 mm).
Display
ESC
SET
LIT
H
L
RUN
RUN
Display value
(mm)
Point 2
4.0
3.8
-3.0
-2.8
-3.0
Point 1
Setting the Output Conditions
Description
"OK/CAN" is displayed on the sub-display.
Press
key to execute the scaling or
ZERO/ SET
ZERO/
SET
to cancel.
RUN
RUN
Press
key for two seconds to enter the RUN
FUN
FUN
L
TEACH
TEACH
mode.