Vega VEGAPULS 62 Operating Instructions Manual page 56

4...20 ma/hart two-wire
Hide thumbs Also See for VEGAPULS 62:
Table of Contents

Advertisement

Input variable
Measured variable
Fig. 30: Data of the input variable
Reference plane
1
2
Measured variable, max. measuring range
Antenna length
3
4
Useful measuring range
Standard electronics
Max. measuring range
Recommended measuring range
Antenna ø 40 mm (1.575 in)
-
Antenna ø 48 mm (1.89 in)
-
Antenna ø 75 mm (2.953 in), ø 95 mm
-
(3.74 in), parabolic antenna
Electronics with increased sensitivity
Max. measuring range
Recommended measuring range
Antenna ø 40 mm (1.575 in)
-
Antenna ø 48 mm (1.89 in)
-
Antenna ø 75 mm (2.953 in)
-
Antenna ø 95 mm (3.74 in)
-
Parabolic antenna
-
Output variable
Output signal
VEGAPULS 62 • 4 ... 20 mA/HART two-wire
The measured quantity is the distance between
process fitting of the sensor and product surface.
The reference plane is the seal surface on the
hexagon or the lower side of the flange.
3
2
4
35 m (114.83 ft)
up to 15 m (49.21 ft)
up to 20 m (65.62 ft)
up to 35 m (114.83 ft)
75 m (246.1 ft)
up to 15 m (49.21 ft)
up to 20 m (65.62 ft)
up to 40 m (131.2 ft)
up to 50 m (164 ft)
up to 75 m (246.1 ft)
4 ... 20 mA/HART
11 Supplement
1
55

Advertisement