Function check
Out of specification
Maintenance (mainte-
nance requirement)
VEGAPULS 62 • 4 ... 20 mA/HART two-wire
The following table shows the codes and text messages of the status
message "Function check" and provides information on causes as well
as corrective measures.
Code
Cause
Text mes-
sage
C700
l Simulation active
Simulation
The following table shows the codes and text messages of the status
message "Out of specification" and gives instructions for the reason
and removal.
Code
Cause
Text mes-
sage
S600
l Temperature of the elect-
ronics in the not specified
Unpermissible
section
electronics
temperature
The following table shows the codes and text messages of the status
message "Maintenance" and provides information on causes as well
as corrective measures.
Code
Cause
Text mes-
sage
M500
l With the reset to delivery
status, the data could not
Error with the
be restored
reset delivery
status
M501
l Hardware error EEPROM
Error in the
non-active li-
nearization
table
M502
l Hardware error EEPROM
Error in the
diagnosis me-
mory
9 Diagnosis and service
Removal
l Finish simulation
l Wait for the automatic end
after 60 mins.
Removal
l Check ambient temperatu-
re
l Isolate electronics
l Use instrument with higher
temperature range
Removal
l Repeat reset
l Load XML file with sensor
data into the sensor
l Exchange the electronics
l Send instrument for repair
l Exchange the electronics
l Send instrument for repair