5.2
Carriage Junction Box Calibration Mode
The load cell junction box must be placed in the calibration mode:
1. Remove the scale cover.
2. Remove the one screw from the plate at the top of the junction box.
3. Swivel plate away from switch opening.
4. Place switch in the calibration position, away from coiled cable connection or toward the right-hand side of junction box
when standing in front of the scale (load cell #2).
Calibration switch
Calibration switch cover
plate
5.3
Standard Calibration
1. Place the junction box in setup mode; see
and select the Setup tab.
2. Press
3. Select Calibration.
ON
Figure 5-3. Junction Box Calibration Switch Location
Section
5.2.
Figure 5-4. Setup Selection
Figure 5-5. Calibration Selection
© Rice Lake Weighing Systems ● All Rights Reserved
Direction of switch for
calibration
OFF
Calibration
Front of forklift
39