CLS-M Installation/Service Manual
4.7
Software Version
Press Get Current Software Version to view the version of software installed on the junction box and the part number of the
junction box installed in the scale.
4.8
Default Configuration
Use the default configuration tab to set configuration and calibration values to factory default.
Use this feature if instructed by the factory or prior to replacing a junction box in the field that appears to be working properly
with Revolution, but not with the mobile computer or hand-held device.
Using the default configuration feature requires a new Zero Angle and Standard Calibration in the setup menu.
IMPORTANT
Use the following steps to set up the default configuration:
1. Place the calibration switch into the open position.
2. Press Default Configuration.
3. Press Yes to continue. A default configuration confirmation displays.
4. Press OK to exit.
www.RiceLake.com
34
Visit our website
Figure 4-37. Software Version Tab
Figure 4-38. Default Configuration Tab
Figure 4-39. Default Configuration – Warning