Adjusting the footplate angle (not illustrated)
1) Loosen and remove the Allen head screws/counter nuts
at the top end of the footrest receiver below the seat bot
tom.
2) Hold the footplate at the desired angle.
3) Tighten the Allen head screws/counter nuts.
6.8 Adjusting the stand-up function
Due to the complexity of the stand-up function, programming
changes or mechanical changes to limit switches and actu
ators may only be made by qualified personnel with special
training.
Further information is available in the training documents or
service instructions.
Adjusting the angle of the stand-up function
This stand-up function setting may also be changed by qual
ified personnel without specialised training.
The possible angle of the stand-up function can be limited
by adjusting the position of the top limit switch for the stand
ing actuator.
1) Move the seat into the standing position until the corres
ponding screw is freely accessible.
2) Loosen the Allen head screw on the actuator (see
fig. 11, item 1).
3) Slide the ring on the scale up to increase the standing
angle or down to decrease the angle.
C1000 SF
4) Retighten the Allen head screw using only minimal
torque.
INFORMATION: The line spacing on the scale is 10
mm. Move the ring by 5 mm at a time and test the
effect. Gradually approach the desired position.
Fig. 11
Adjusting the stand-up function actuator
6.9 Adjusting the belts
6.9.1 Safety instructions
CAUTION
Incorrect approach to the adjustment process
Injuries, malpositions, illness of the user due to adjustment
errors
Settings