Programming, Calibration & Test Universal Bill Acceptor (UBA-1x-SS & UBA-x4-SS/SU)
Figure 6-23 MAG HEAD Adjustment Dialog
Screen
Figure 6-24 MAG HEAD Adjustment Successful
Screen
13. Insert the Mag Head Test Board into the
UBA. Set the Mag Head Test Board so the
middle line is located just above the Roller
located on the Lower Tray, then close the
guide firmly and Mouse-click on the
Screen Button.
14. Move the Mag Head Test Board back and
forth, in and out slightly, to find the peak
value. Find the position where the
value enters a range within
the peak value, then press the
NOTE: The average peak value range is
approximately 20P to approximately 35P.
Use this value only for reference and
make sure to find the true peak value
using your particular Acceptor Unit.
15. When the MAG HEAD Adjustment is com-
plete, the
"ADJUSTMENT SUCCESSFUL!!"
Screen shown in Figure 6-24 will appear.
16. Mouse-click on the
ton again, and this calibration process will be
complete.
Calibration Error Codes that may occur dur-
ing these procedures are listed in Table 6-1.
JCM Part No. 960-000097R_Rev. 3
OK
"GAIN1"
in relation to
-5P
Bar.
"Space"
Screen But-
OK

Table 6-1 Calibration Error Table

Error
Displayed
Code
Message
Gain Error
4-A
(Value over 4.3V)
4-B
Adjustment Error
4-C
Black Level Error
4-E
Gain Error
Front/Back/Pbin/
4-G
Width Level Error
6-A
6-B
Offset Error
6-C
Adjustment Error Magnetic Sensor Adjustment
MAG
Adjustment Error
Under 0.74V
Gain Max Limit
Over Error
Bar Gain Max
Limit Over Error
UV Gain Max
Limit Over Error
No Code
A/D Data Level
Error
Pbin/Width D/A
Error
Motor Speed
Error
EEPROM Write
Error

UBA-USB Tool Suite Overview

The UBA-USB Tool Suite is a Software Appli-
cation allowing Software updating, performance
data retrieval and performing Calibration of the
UBA Unit using the USB Port of the Acceptor to
download.
6 - 7
Section 6
Description/Cause
Light receiving Adjustment
Error. Check for dirty or wrong
Calibration Paper use.
Sensor Light Quantity
Adjustment Error. Replace
either the Upper or Lower
Sensor PCB
Sensor Light Quantity
Adjustment Error. Ensure that
the Black Reference Paper is
properly inserted.
Light Receiving Adjustment
Error. Clean Sensors. Replace
Upper or Lower Sensor PCB.
Triggering Sensor Light
Receiving Error. Clean the
Sensor. Wrong Calibration
Paper being used during UV
Sensor Adjustment. Replace
the Exit Sensor.
NOTE: The UBA must in
installed in a Frame during
Calibration.
Light Receiver Circuit
Abnormality. Clean Sensors.
Replace either the Upper or
Lower Sensor PCB.
Error. Replace the Upper
Sensor PCB.
Sensor Abnormality. White
Reference Paper not inserted
correctly.
Sensor Abnormality. For UBA
1x, Replace the Upper Sensor
PCB. For UBA 24, Replace the
Lower Sensor PCB
Sensor Abmormality. Clean UV
Sensor and White Reference
Block. Ensure UV Reference
Paper is insert with its Label
up, covering the White
Reference block.
Light Receiving Level Error.
Ensure the White Reference
Paper is inserted completely
and correctly placed
Triggering Sensor Adjustment
Error. Clean PB In-Sensor or
Reference Paper is not
inserted correctly
Transport Motor Speed Error
Adjustment Value Writing Error.
Replace the Processor PCB.
© 2010, JCM American, Corporation