5.5.12

Loading factory default values (on the 3102/3105)

Screen display:
It may, occasionally, be necessary to re-set the transmitter parameters to factory default values,
particularly if the data already changed is in question.
Note
Re-loading factory defaults overwrites all parameters and all site entered data will be lost.
To ensure that this operation is not initiated by accident, a specific integral button sequence is
necessary to load factory defaults.
Note
This menu option is in the engineering menu "Eng". See
the 3102/3105)" on page 62
To load factory default values
1.
2.
3.
4.
5.
6.
Service and Troubleshooting
Ld.dEF
or see
After entering the "Eng" menu (see the note above), press the green button
repeatedly until "Ld.dEF" is indicated.
Press the blue button
to enter the load defaults "Ld.dEF" menu and display "LOAd".
Press and hold the blue button
Press the blue button
again to acknowledge the message and stop the message
flashing.
If you do not want to continue, press the blue button
green button
to get to the next menu option. Otherwise, continue with step (6).
To load factory defaults, press and hold both the blue
two seconds.
The screen flashes "b.units" and reloads all factory default values. The display then
changes to the PV display.
Section 5: Service and Troubleshooting
"Accessing the engineering menu (on
Figure C-5 on page 106
for a map of the menu structure.
for two seconds to flash the screen message "SUrE".
February 2015

to abort and then press the

and red
buttons together for
73