Ultrasonic level and flow transmitter
Hide thumbs Also See for Rosemount 3107:

Advertisement

Quick Start Guide
00825-0200-4840, Rev BC
August 2019
Rosemount
3107 Level and 3108 Flow
Transmitters
Ultrasonic

Advertisement

loading

Summary of Contents for Emerson Rosemount 3107

  • Page 1 Quick Start Guide 00825-0200-4840, Rev BC August 2019 ™ Rosemount 3107 Level and 3108 Flow Transmitters Ultrasonic...
  • Page 2: Table Of Contents

    Quick Start Guide August 2019 Contents About this guide...........................3 Overview of Rosemount 3107 and 3108..................6 Considerations before you install....................9 Installation..........................11 Connecting the transmitter......................16 Configuration..........................18 Product Certifications.........................33 Rosemount Ultrasonic 3107 Level and 3108 Flow Transmitters...
  • Page 3: About This Guide

    August 2019 Quick Start Guide About this guide This installation guide provides basic guidelines for the Rosemount 3107 Level Transmitter and Rosemount 3108 Flow Transmitter. It does not provide instructions for detailed configuration, diagnostics, maintenance, service, troubleshooting, or installations. Refer to the Rosemount 3107 and...
  • Page 4 Failure to follow these installation guidelines could result in death or serious injury. • The Rosemount 3107 and Rosemount 3108 are ultrasonic transmitters. They must be installed, connected, commissioned, operated, and maintained by suitably qualified personnel only, observing any national and local requirements that may apply.
  • Page 5 August 2019 Quick Start Guide WARNING Physical access Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could be intentional or unintentional and needs to be protected against. Physical security is an important part of any security program and fundamental to protecting your system.
  • Page 6: Overview Of Rosemount 3107 And 3108

    Quick Start Guide August 2019 Overview of Rosemount 3107 and 3108 The Rosemount 3107 and 3108 are sealed 4–20 mA loop-powered liquid level transmitters, specifically designed for use in waste water and effluent treatment plant on aqueous applications. These rugged UPVC transmitters are certified Intrinsically Safe for use in Zone 0 areas, and factory fitted with up to 165 ft.
  • Page 7 August 2019 Quick Start Guide Figure 2-1: Typical Application A. Rosemount 3108 Flow Transmitter B. Rosemount 3490 Series Control Unit C. 4–20 mA signal output D. Relay E. Pump F. Transmitter Bottom Reference G. 4–20 mA and HART signal input Components of the transmitter The transmitter has a housing containing advanced electronics to generate ultrasonic pulses, process the resultant signals, and provide a 4–20 mA and...
  • Page 8 Quick Start Guide August 2019 Figure 2-2: Transmitter Components A. Mounting bracket B. Two-core cable C. 1-in. mounting thread D. UPVC wetted parts E. Remote temperature sensor (Rosemount 3108 only) Rosemount Ultrasonic 3107 Level and 3108 Flow Transmitters...
  • Page 9: Considerations Before You Install

    It is not advisable to mount the transmitter in close proximity to a source of electrical noise such as a variable-speed drive, or other high-powered electrical device. Environmental The Rosemount 3107 and 3108 ultrasonic transmitters are Intrinsically Safe (IS) approved for hazardous area installations. Quick Start Guide...
  • Page 10 Quick Start Guide August 2019 • The Rosemount 3107 is designed for open or closed tank installation. It is weatherproof and protected against the ingress of dust • The Rosemount 3108 is designed for open channel flow measurement. It is weatherproof and protected against the ingress of dust •...
  • Page 11: Installation

    August 2019 Quick Start Guide Installation • Mount the transmitter above the liquid using the 1-in. threaded connection provided, but no closer than 13.8 in. (0,35 m) to the surface. The transmitter does not detect any liquid surface closer than 12 in. (0,3 m) to the transmitter face.
  • Page 12 Rosemount 3490 Series Control Unit or a Field Communicator. Flange mounting The instrument (accessory) flanges supplied by Emerson are manufactured from PVC and are a full face design. Care must be taken when installing to a raised face mating flange on the tank or vessel to prevent distortion of the PVC flange by overtightening the bolts.
  • Page 13 August 2019 Quick Start Guide Figure 4-2: Recommended Mounting Position A. Transmitter is mounted vertically (maximum deviation of 3°). B. 6° beam half angle. C. 1.3 in./ft. (11 cm/m). Minimum of 12 in. (0.3 m). Mounting from a conduit The transmitters can be mounted from a conduit using the optional adaptor accessory.
  • Page 14 Quick Start Guide August 2019 Figure 4-3: Choosing the Height Position Above a Flow A. Transmitter front face B. Hmax C. Transmitter bottom reference = Hmax + 12.2 in. (300 mm) + 2 in. (50 Figure 4-4: Bottom Reference of a Flume or Weir A.
  • Page 15 If the flow structure permits, mount the transmitter within the flow channel or chamber. Note For some installations, the use of a calibration device is mandatory. Emerson offers the Rosemount Head Verification Device (HVD) for this purpose. See theQuick Installation Guide for more information.
  • Page 16: Connecting The Transmitter

    2 or separate extra-low voltage (SELV) source. Other devices may reset if connecting the transmitter to a multi-drop system while the loop is powered. De-energize loop to avoid devices being reset. The Rosemount 3107 and 3108 are two-wire, loop-powered transmitters accepting external power supplies as follows: •...
  • Page 17 August 2019 Quick Start Guide Figure 5-1: Wiring A. Remote temperature sensor (Rosemount 3108 only) B. Black: 0 Vdc C. Red: 12 to 40 Vdc (non-hazardous area), 12 to 30 Vdc from protective barrier (hazardous area) D. Non-hazardous area: connect cable screen to standard ground (Earth) or hazardous area: connect cable screen to I.S.
  • Page 18: Configuration

    Rosemount 3490 Series Control Unit. The parameters in this section are sufficient for a basic level, contents (volume), or open channel flow application. For a more advanced application, refer to the Rosemount 3107 Level and 3108 Flow Transmitters Reference Manual.
  • Page 19 August 2019 Quick Start Guide Procedure 1. From the Main Menu screen, select SETUP. 2. Select the transmitter (e.g. “Tx1: 3107”). 3. Select SYSTEM, and then select Base Units. 4. Select new base units. Base Units metric Esc=Quit =Edit Note To get the same base units on the control unit, switch the power off and then on again.
  • Page 20 Quick Start Guide August 2019 Bottom Reference P010 40.000 ft 40.000 ft Lock @ & Shift á ü HELP ENTER Rosemount Ultrasonic 3107 Level and 3108 Flow Transmitters...
  • Page 21 August 2019 Quick Start Guide 6.2.2 Viewing of changing P010 (Rosemount 3490 Series Control Unit) To view or change the bottom reference: Procedure 1. From the Main Menu screen, select SETUP. 2. Select the transmitter (e.g. “Tx1: 3107”). 3. Select DUTY, and then select Bottom Ref. 4.
  • Page 22 Quick Start Guide August 2019 6. Press “SEND” to update the transmitter. Note If the HART PV has no units, select and confirm the “None”, “Unknown”, or “Not Used” option as appropriate for the HART Master Device (host). 6.3.2 Viewing or changing P012 (Rosemount 3490 Series Control Unit) To view or change the existing setting: Procedure 1.
  • Page 23 August 2019 Quick Start Guide 5. Select a new profile, and then press “Enter” to save the selection. 6. Press “SEND” to update the transmitter. The selected profile can be viewed at Fast Key sequence 2, 2, 3, 3. Note When on-screen messages appear, take action if needed and press “OK”.
  • Page 24 Figure 6-1: Volume from a Cylinder/Sphere P013 = Full Volume of Ideal Cylindrical or Spherical Tank of Constant Diameter P014 A. Rosemount 3107 or 3108 B. Rosemount 3490 Series Control Unit C. 4–20mA/HART D. Bottom reference (P010) E.
  • Page 25 August 2019 Quick Start Guide 3. Select 3: Profiling. 4. Select 4: Scale Factor P013 or 4: k-factor P013, depending on the Non-Linear Profile selected. 5. Input the factor, and press “ENTER” to save it. Scale Factor P013 Lock @ & Shift á...
  • Page 26 Quick Start Guide August 2019 When the PV is a volume measurement from a regular-shaped tank (e.g. square or rectangular), this parameter is not used. When the PV is a volume measurement from an irregular-shaped tank, use this parameter to enter the maximum height. See also Procedure for P011=“Special Plot”...
  • Page 27 August 2019 Quick Start Guide 4. Select Profile Height. Profile Height P014 1.0000ft Esc=Quit =Edit 5. Follow the on-screen instructions to edit and save the new setting. (Press the Enter ( ) key if prompted to change the mode to “off- line”).
  • Page 28 Quick Start Guide August 2019 can be changed only if the “Special Plotted” profile has been selected. 6.7.2 Viewing or changing P030-P039 (Rosemount 3490 Series Control Unit) To view or change a profile point: Procedure 1. From the Main Menu screen, select SETUP. 2.
  • Page 29 August 2019 Quick Start Guide Figure 6-2: Volume or Flow from 2-stage Weir A. Rosemount 3107 or 3108 B. 4–20 mA/HART C. Rosemount 3490 Series control unit D. Rectangular weir E. V-notch weir F. Profile height (P014) G. Bottom reference (P010)
  • Page 30 Quick Start Guide August 2019 Figure 6-3: Graph of PV versus Height P039 P038 P037 P036 P035 P034 P033 P032 P031 D D D P030 20% 30% 40% 50% 60% 70% 80% 90% 100% A. (Entered percentages relating output PV to maximum PV) B.
  • Page 31 Level offset (P069) User Preferred SRP (UPSRP) Level SV (D901) Distance offset (P060) Distance (D910) Sensor Reference Point (SRP) Liquid Level 4–20 mA/HART Distance TV (D902) Rosemount 3107 or 3108 Lower blanking (P063) Rosemount 3490 Series control unit Quick Start Guide...
  • Page 32 Quick Start Guide August 2019 Rosemount 3107 and 3108 parameters on the Rosemount 3490 The process value (e.g. liquid level) is indicated in the HART Primary Variable (D900), which drives the 4–20 mA output signal. Parameter Fast key 3490 Series menu navigation...
  • Page 33: Product Certifications

    Emerson.com/Rosemount. ATEX directive (94/9/EC) Emerson complies with the ATEX directive. Pressure Equipment Directive (PED) (97/23/EC) The Rosemount 3107 and 3108 are outside the scope of PED directive. Electro Magnetic Compatibility (EMC) (2004/108/EC) EN 61326-1:2006 MCERTS certification...
  • Page 34 Quick Start Guide August 2019 Special Condition for Safe Use (X): 1. To protect against UV exposure when installed outdoors, the transmitter shall be installed in accordance with the section Environmental. I6 Canadian Standards Association (CSA) Certificate Number: 02 CSA 1352094 X Ex ia IIC Markings: Intrinsically Safe when installed with certified...
  • Page 35 August 2019 Quick Start Guide The following instructions apply to equipment covered by certificate number Sira 09ATEX2299X: 1. The equipment may be used with flammable gases and vapors with apparatus groups IIA, IIB, and IIC, and with temperature classes T1, T2, T3, T4, T5, and T6.
  • Page 36 Quick Start Guide August 2019 b. The equipment must only be cleaned with a damp cloth. c. Do not mount the 3107/3108 on a structure that is subject to vibration, or in a position where damage may be caused by impact or thermal stress.
  • Page 37 August 2019 Quick Start Guide EU Declaration of Conformity Figure 7-1: EU Declaration of Conformity Quick Start Guide...
  • Page 38 Quick Start Guide August 2019 Rosemount Ultrasonic 3107 Level and 3108 Flow Transmitters...
  • Page 39 August 2019 Quick Start Guide China RoHS Rosemount 3107/8 China RoHS List of Rosemount 3107/8 Parts with China RoHS Concentration above MCVs / Hazardous Substances Hexavalent Polybrominated Polybrominated Part Name Lead Mercury Cadmium Chromium biphenyls diphenyl ethers (Pb) (Hg) (Cd)
  • Page 40 The Emerson logo is a Twitter.com/Rosemount_News trademark and service mark of Emerson Electric Facebook.com/Rosemount Co. Rosemount is a mark of one of the Emerson Youtube.com/user/ family of companies. All other marks are the RosemountMeasurement property of their respective owners.