5.2.8
Measured values display and operating procedures
1 : READ PARAMETERS
2 : MEASURED VALUES
3 : SELF-DIAGNOSIS
[ESC]
(Previous
Menu)
MEASURED VALUES *
p PHASE:
T TEMP:
X DENSITY:
5.2.9
Self-diagnosis data display operating procedures
1 : READ PARAMETERS
2 : MEASURED VALUES
3 : SELF-DIAGNOSIS
ST:STATUS :GOOD
[ESC]
(Previous
Menu)
[ESC] RETURN
[DN]
(Pre. menu)
SL:MICROWAVE SIG.LVL
DATA :−53 dBm
[ESC]
(Previous
[ESC] RETURN
Menu)
Move the cursor to "2" with [→] key, and press [SET] key to
select [ 2:MEASURED VALUES]
123.56 N=0
°C
30.00
2.34
%TS
Move the cursor to "3" with [→] key,
and press [SET] key to select [3:SELF-DIAGNOSIS]
Whether the phase m easurement operation is
normally functioning or not can be verified.
[UP]
(Next menu)
The micro wave receive signal level can be
verified. Normally, the level is -45 to -75. (dBm)
[UP]
(Next menu)
41
It is possible to verify the measured phase value (and
the angle rotation N), the measured fluid temperature
value, as well as the density calculated based on them.
[*] in the upper right corner will blink synchronously with
the data updating thus making it possible to verify that
the data updating is definitely taking place.