Measured Values Display And Operating Procedures; Self-Diagnosis Data Display Operating Procedures - Toshiba LQ500 Operation Manual

Density meter
Hide thumbs Also See for LQ500:
Table of Contents

Advertisement

5.2.8

Measured values display and operating procedures

1 : READ PARAMETERS
2 : MEASURED VALUES
3 : SELF-DIAGNOSIS
[ESC]
(Previous
Menu)
MEASURED VALUES *
p PHASE:
T TEMP:
X DENSITY:
5.2.9

Self-diagnosis data display operating procedures

1 : READ PARAMETERS
2 : MEASURED VALUES
3 : SELF-DIAGNOSIS
ST:STATUS :GOOD
[ESC]
(Previous
Menu)
[ESC] RETURN
[DN]
(Pre. menu)
SL:MICROWAVE SIG.LVL
DATA :−53 dBm
[ESC]
(Previous
[ESC] RETURN
Menu)
Move the cursor to "2" with [→] key, and press [SET] key to
select [ 2:MEASURED VALUES]
123.56 N=0
°C
30.00
2.34
%TS
Move the cursor to "3" with [→] key,
and press [SET] key to select [3:SELF-DIAGNOSIS]
Whether the phase m easurement operation is
normally functioning or not can be verified.
[UP]
(Next menu)
The micro wave receive signal level can be
verified. Normally, the level is -45 to -75. (dBm)
[UP]
(Next menu)
 41 
It is possible to verify the measured phase value (and
the angle rotation N), the measured fluid temperature
value, as well as the density calculated based on them.
[*] in the upper right corner will blink synchronously with
the data updating thus making it possible to verify that
the data updating is definitely taking place.