[DN]
(Pre. menu)
a: DENSITY LINE SLOPE
DATA: 0.0840
[ESC]
(Previous
[ESC] RETURN
Menu)
[DN]
(Pre. menu)
b: DENSITY INTERCEPT
DATA: 0.0
[ESC]
(Previous
[ESC] RETURN
Menu)
[DN]
(Pre. menu)
ot: DENSTY TEST OUT
DATA: 1.5
[ESC]
(Previous
[ESC] RETURN
Menu)
[DN]
(Pre. menu)
dt: DELAYED SYNC.TIME
DATA: 0.5
[ESC]
(Previous
[ESC] RETURN
Menu)
[DN]
(Pre. menu)
zp:ZERO POINT PHASE
DATA: 80.00
[ESC]
[ESC] RETURN
(Previous
Menu)
[DN]
(Pre. menu)
zT:ZERO WATER TEMP.
DATA: 25.00
[ESC]
(Previous
[ESC] RETURN
Menu)
[UP]
(Next menu)
The set value of "density line slope" of the
arithmetic expression for calculating the density
from the phase measurement data, etc. can be
verified.
[UP]
(Next menu)
The set value of "density intercept" of the arithmetic
expression for calculating the density from the
phase measurement d ata, etc. can be verified.
Normally, this is set to zero.
[UP]
(Next menu)
Upon getting into [2: SETTING MENU], the current
output and the LED density display are switched to
simulated values. In this screen, the set value of the
simulated value (unit: %TS) can be verified.
[UP]
(Next menu)
In external synchronized operations, the set value
of the delayed time (unit: minute) from when the
external contact input is turned ON until the
measurement starts can be verified. For details,
refer to Section 6.7.
[UP]
(Next menu)
The phase at the time of zero calibration can be
verified.
[UP]
(Next menu)
The fluid temperature at the time of zero calibration can
be verified.
 39 