5.2.10

Parameter setting display and operating procedures

5:SET PARAMETERS
6:ZERO CALIBRATION
7:SPAN CALIBRATION
8:ANGLE ROTATION
UR:UPPER RANGE
DATA: 3.0
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN.
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
LR:LOWER RANGE
DATA: 0.0
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
a:DENSITY LINE SLOPE
DATA: 0.0840
[SET] CHANGE
[ESC]
[ESC] RETURN
(Previous
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
b:DENSITY INTERCEPT
DATA: 0.00
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
ot:DENSTY TEST OUT.
DATA: 1.5
[SET] CHANGE
[ESC]
[ESC] RETURN
(Previous
Menu)
[UP]
(Next menu)
Move the cursor to "5" with [→] key, and press [SET] key to select[ 5:SET PARAMETERS]
[SET](Set)
[SET](Fix)
[ESC](Cancel)
[SET](Set)
[SET](Fix)
[ESC](Cancel)
[SET](Set)
[SET](Fix)
[ESC](Cancel)
The "density line slope" of the arithmetic expression for
calculating the density from the phase measurement data, etc. is
set. Before shipping, this is s et in accordance with the aperture,
based on Table 5.1.2. Normally, do not change the setting.
[SET](Set)
[SET](Fix)
[ESC](Cancel)
The "density intercept" of the arithmetic expression for calculating
the density from the phase measurement data, etc. is set. In the
event that the zero point is out of position, it is corrected through
zero calibration; it is also possible to shift the zero point by
changing this coefficient. Normally, set the value to zero.
[SET](Set)
[SET](Fix)
[ESC](Cancel)
UR:UPPER RANGE
RANGE :1.0-99.9 %TS
DATA : 0.30
[SET] SET,[ESC] CANCEL
The upper density measurement range
(whose current output is 20mA) is set.
LR:LOWER RANGE
RANGE :0.0-99.5 %TS
DATA :00.0
[SET] SET,[ESC] CANCEL
The lower density measurement range
(whose current output is 4mA) is set.
a:DENSITY LINE SLOPE
RANGE :−0.2000−0.2000
DATA : 0.0840
[SET] SET,[ESC] CANCEL
b:DENSITY INTERCEPT
RANGE :−99.99−99.99
DATA : 00.00
[SET] SET,[ESC] CANCEL
ot:DENSTY TEST OUT.
RANGE :0.0-99.9 %TS
DATA : 01.5
[SET] SET,[ESC] CANCEL
When [2: SETTING MENU] is selected, the density output
signal and the density display are hold in the simulated values
that have been set. In this screen, the simulated values are
set in terms of "%TS".