5.2.8

Measured values display and operating procedures

:READ PARAMETERS
1
2 :MEASURED VALUES
3 :SELF-DIAGNOSIS
[ESC]
(Previous
Menu)
p PHASE: 123.56 N=0
T LIQ_TMP:  30.00 ℃
A AT_TMP :  30.00 ℃
*X DENSITY: 2.34 %TS:
5.2.9

Self-diagnosis data display operating procedures

:READ PARAMETERS
1
2 :MEASURED VALUES
3 :SELF-DIAGNOSIS
ST:STATUS :GOOD
[ESC]
(Previous
[ESC] RETURN
Menu)
[UP]
SL:MICROWAVE SIG.LVL
DATA :−53
[ESC]
(Previous
[ESC] RETURN
Menu)
[UP]
F:MICROWAVE COEF.
DATA : 1910
[ESC]
(Previous
[ESC] RETURN
Menu)
[UP]
Move the cursor to "2" with [→] key, and press [SET] key to
select [ 2:MEASURED VALUES]
Move the cursor to "3" with [→] key,
and press [SET] key to select [3:SELF-DIAGNOSIS]
[DN]
[DN]
 44 
It is possible to verify the measured phase value (and
the angle rotation N), the measured liquid temperature
value, the measured atmospheric temperature value, as
well as the density calculated based on them. [*] at the
left end of the bottom line will blink synchronously with
the data updating thus making it possible to verify that
the data updating is definitely taking place.
Whether the phase measurement operation is
normally functioning or not can be verified.
The micro wave receive signal level can be
verified. Normally, the level is -45 to -75.
The micro wave signal constant can be verified.
Normally, the value is 1825 to 1975.