Other Menus Display And Operating Procedures - Toshiba LQ500 Operation Manual

Density meter
Hide thumbs Also See for LQ500:
Table of Contents

Advertisement

6 F 8 A 0 5 2 1
5.2.16

Other menus display and operating procedures

9:LINEARIZ/CNDUCTVTY
10:ADDITIVES CORRECT
11:OTHERS
ho:4-20mA IN [EXT]
DATA :4mA
[SET] CHANGE
[ESC]
[ESC] RETURN
(Previous
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
DI:C CHANGE. ON/OFF
DATA :OFF
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
C2:DensityMultiplier
DATA: 1.000
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
C3:DensityMultiplier
DATA: 1.000
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[UP]
(Next menu)
Move the cursor to "11" with [→] key, and press [SET] key to select
[ 11:OTHERS]
[SET](Set)
[SET](Fix)
[ESC](Cancel)
The simulated output at contact OFF in the event of external
operation can be selected. Each time the[UP] or [DN] key is pressed, the mode
display is switched around among LAST VALUE (immediately preceding value),
4mA and TEST OUTPUT (simulated output 0.0 to 99.9%TS when in setting mode).
[SET] is pressed with the mode display to select.
[SET](Set)
[SET](Fix)
[ESC](Cancel)
Up to four density multipliers can be switched around by the external voltage
signal (DI) of two points. This is used for switching the brand, etc. On this
menu, whether to perform the density multipliers switching (ON) or not (OFF)
is selected. Each time the [UP] or [DN] key is pressed, ON/OFF are mutually
alternated.
[SET](Set)
[SET](Fix)
[ESC](Cancel)
In the event of performing the density multiplier switching by means of the
external voltage signal (DI) of two points, the second density multiplier is set. In
the same manner, the following 3rd and 4th density multipliers are set. The first
density multiplier is performed with the menu for span calibration.
[SET](Set)
[SET](Fix)
[ESC](Cancel)
 54 
ho:4-20mA IN [EXT]
RANGE :LAST/4mA/TEST
DATA: 4mA
[SET] SET,[ESC] CANCEL
synchronized
DI:C CHANGE. ON/OFF
RANGE :ON/OFF
DATA: OFF
[SET] SET,[ESC] CANCEL
C2:DensityMultiplier
RANGE :0.000-9.999
DATA: 1.000
[SET] SET,[ESC] CANCEL
C3:DensityMultiplier
RANGE :0.000-9.999
DATA: 1.000
[SET] SET,[ESC] CANCEL