Density meter
Hide thumbs Also See for LQ500:
Table of Contents

Advertisement

[DN]
(Pre. menu)
zG:ZERO RF DATA
DATA: 50.00
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
ma:AVERAGING TIMES
DATA: 1
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
dx:CHANGE RATE LIMIT
DATA: 0.00
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
HL:LIMIT TIMES
DATA: 0
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
UR:UPPER RANGE
DATA: 3.0
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[NOTE]:
For all menus, when data outside the allowed range is
input, the error message is displayed, and the wrong data is
refused. Press any key to return, and input adequate data.
[SET](Set)
[SET](Fix)
[ESC](Cancel)
The zero-point RF data can be set in this menu through
manual input. Normally, the zero-point RF constant is
automatically set when implementing the zero calibration.
[SET](Set)
[SET](Fix)
[ESC](Cancel)
The moving average times are set. If the output averaged for
density control, etc. is needed, the value from 5 to about 10 is
set.
[SET](Set)
[SET](Fix)
[ESC](Cancel)
If there is a signal of sharp variation, the change-rate limit function
excludes this signal to restrict a sudden output variation. In this
menu, the permissible variation width is set in terms of "%TS".
[SET](Set)
[SET](Fix)
[ESC](Cancel)
The limit times in the change-rate limit function are set. If the
variation width exceeds dx(%TS) that was set in the previous
menu, the signal is excluded for up to the HL times that are set in
this menu. At HL=0, the change-rate limit function does not work.
[SET](Set)
[DN]
 45 
zG:ZERO RF DATA
RANGE :0.00-100.00 (C
DATA: 050.00
[SET] SET,[ESC] CANCEL
ma:AVERAGING TIMES
RANGE : 1-99
DATA : 01
[SET] SET,[ESC] CANCEL
dx:CHANGE RATE LIMIT
RANGE :0.00-9.99 °
DATA: 0.00
[SET] SET,[ESC] CANCEL
HL:LIMIT TIMES
RANGE :0-99
DATA: 00
[SET] SET,[ESC] CANCEL
VALUE OUT OF RANGE
UR:UPPER RANGE
PRESS ANY KEY TO
RETURN