Toshiba LQ500 Operation Manual page 44

Density meter
Hide thumbs Also See for LQ500:
Table of Contents

Advertisement

6 F 8 A 0 5 2 1
[DN]
(Pre. menu)
F:MICROWAVE COEF.
DATA : 1910
[ESC]
(Previous
[ESC] RETURN
Menu)
[DN]
(Pre. menu)
G:RF DATA.
DATA : 50.00
[ESC]
[ESC] RETURN
(Previous
Menu)
[DN]
(Pre. menu)
J:+5V POWER SUPPLY
DATA :5.00
[ESC]
[ESC] RETURN
(Previous
Menu)
[DN]
(Pre. menu)
pd:REF PHASE
DATA: 0.0
[ESC]
(Previous
[ESC] RETURN
Menu)
[DN]
(Pre. menu)
Mc:MEMORY CHECK:GOOD
EPROM :GOOD
RAM
[ESC]
EEPROM :GOOD
(Previous
Menu)
[DN]
(Pre. menu)
ST:STATUS :GOOD
[ESC]
(Previous
[ESC] : RETURN
Menu)
[UP]
(Next menu)
[UP]
(Next menu)
[UP]
(Next menu)
[UP]
(Next menu)
:GOOD
[UP]
(Next menu)
 42 
The micro wave signal constant can be verified.
Normally, the value is 1825 to 1975.
The converter's RF data (concerning the micro
wave phase measurement) can be verified.
Normally, the value is 10 to 70.
The voltage of the 5-volt power supply can be
checked.
The soundness of the phase measurement unit
can be verified.
The soundness of various memory units can be
verified.