6 F 8 A 0 5 2 1
5.2.14

Linearize/conductity correction display and operating procedures

9:LINEARIZ/CNDUCTVTY
10:ADDITIVES CORRECT
11:OTHERS
LA:DENSITY A
DATA :0.60
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
LB:DENSITY B
DATA: 1.00
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
K1:COEF.
DATA: 1.00
[SET] CHANGE
[ESC]
[ESC] RETURN
(Previous
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
K2:COEF.
DATA: 1.00
[SET] CHANGE
[ESC]
[ESC] RETURN
(Previous
Menu)
[DN]
[UP]
(Pre. menu)
(Next menu)
K3:COEF.
DATA: 1.00
[SET] CHANGE
[ESC]
[ESC] RETURN
(Previous
Menu)
[UP]
(Next menu)
[DN]
(Pre. menu)
r:CNDUCTVTY. COEF.
RANGE : 0.00
Move the cursor to "9" with [→] key, and press [SET] key to
select [ 9:LINEARIZ/CNDUCTVTY]
[SET](Set)
[SET](Fix)
[ESC](Cancel)
[SET](Set)
[SET](Fix)
[ESC](Cancel)
[SET](Set)
[SET](Fix)
[ESC](Cancel)
[SET](Set)
[SET](Fix)
[ESC](Cancel)
[SET](Set)
[SET](Fix)
[ESC](Cancel)
 48 
[SET](Set)
[SET](Fix)
LA:DENSITY A
RANGE :0.00-99.99
DATA :00.60
[SET] SET,[ESC] CANCEL
In the event of three straight lines which closely analogous
each other, the density of the crossover point A of the 2nd
and 3rd lines is set.
LB:DENSITY B
RANGE :0.00-99.99
DATA : 01.00
[SET] SET,[ESC] CANCEL
In the event of three straight lines which closely analogous
each other, the density of the crossover point B of the 2nd
and 3rd lines is set.
K1:COEF.
RANGE:0.00-9.99
DATA : 1.00
[SET] SET,[ESC] CANCEL
In the event of three straight lines which closely analogous
each other, the inclination K1 of the 1st line is set.
K2:COEF.
RANGE:0.00-9.99
DATA : 1.00
[SET] SET,[ESC] CANCEL
In the event of three straight lines which closely analogous
each other, the inclination K2 of the 2nd line is set.
K3:COEF.
RANGE :0.00-9.99
DATA : 1.00
[SET] SET,[ESC] CANCEL
In the event of three straight lines which closely analogous
each other, the inclination K2 of the 3rd line is set.
r:CNDUCTVTY. COEF.
RANGE :0.00-99.99