Setting The Loop To Manual; Wiring Diagrams; Basic Setup - Emerson Rosemount 3051S ERS Reference Manual

Electronic remote sensor system
Hide thumbs Also See for Rosemount 3051S ERS:
Table of Contents

Advertisement

Configuration
November 2016
Electrical shock could cause death or serious injury.
If the Rosemount
environment and a fault or installation error occurs, high voltage may be present on the transmitter
leads and terminals.
Use extreme caution when making contact with the leads and terminals.
Process leaks could result in death or serious injury.
Install and tighten all four flange bolts before applying pressure.
Do not attempt to loosen or remove flange bolts while the transmitter is in service.
Replacement equipment or spare parts not approved by Emerson
reduce the pressure retaining capabilities of the transmitter and may render the instrument
dangerous.
Use only bolts supplied or sold by Emerson as spare parts.
Improper assembly of manifolds to traditional flange can damage the device.
For the safe assembly of the manifold to the transmitter flange, bolts must break the back plane of
the flange web (i.e. bolt hole) but must not contact the sensor module.
Static electricity can damage sensitive components.
Observe safe handling precautions for static-sensitive components.
3.2.1

Setting the loop to manual

Whenever sending or requesting data that would disrupt the loop or change the output of the
transmitter, set the process application loop to manual. The Field Communicator or AMS Device
Manager will prompt to set the loop to manual when necessary. Acknowledging this prompt does not set
the loop to manual. The prompt is only a reminder; set the loop to manual as a separate operation.
3.3

Wiring diagrams

Connect a Field Communicator or AMS Device Manager using a wiring configuration as shown in
Figure
connected at "PWR/COMM" on the terminal block of the Rosemount 3051S ERS Primary Transmitter,
across the load resistor, or at any termination point in the signal loop.
The Field Communicator or AMS Device Manager will search for a HART
when the connection is made. If the Field Communicator or AMS Device Manager fail to connect, it
indicates that no device was found. If this occurs, refer to
3.4

Basic setup

It is recommended that the following items are verified and configured to ensure the proper
functionality of the ERS System.
26
3051S Electronic Remote Sensor (ERS)
2-9,
Figure
2-10, or
Figure
2-11. The Field Communicator or AMS Device Manager may be
Section 5:
Reference Manual
00809-0100-4804, Rev CB
System is installed in a high-voltage
for use as spare parts could
®
-compatible device and indicate
Troubleshooting.
Configuration

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 3051s

Table of Contents