Table 5-1. Diagnostic Message Troubleshooting
LCD display
Host diagnostic
message
message
CURR
mA Output
SAT
Saturated
DP
System DP Alert
ALERT
FAIL
BOARD
Electronics Error
ERROR
FAIL
P
P
Module Failure
HI
HI
ERROR
FAIL
P
P
Module Failure
LO
LO
ERROR
FAIL
T
P
Module Failure
HI
HI
ERROR
FAIL
T
P
Module Failure
LO
LO
ERROR
P
HI
P
Pressure Alert
HI
ALERT
P
P
Module
HI
HI
COMM
Communication
ERROR
Error
P
P
Pressure Out of
HI
HI
LIMIT
Limits
P
LO
P
Pressure Alert
LO
ALERT
P
P
Module
LO
LO
COMM
Communication
ERROR
Error
P
P
Pressure Out of
LO
LO
LIMIT
Limits
LOOP
mA Output Fixed
TEST
SNSR
Sensor Module
COMM
Missing
ERROR
SNSR
No P
Module
HI
CONFIG
Configuration
ERROR
Present
56
Possible problems
The primary variable has exceeded the
range points defined for the 4–20 mA
analog output signal. The analog output is
fixed at the high or low saturation point
and is not representative of the current
process conditions.
The ERS System is measuring a differential
pressure value that exceeds the configured
upper or lower alert value.
The electronics feature board in the ERS
Primary Unit has malfunctioned.
The P
sensor module has failed.
HI
The P
sensor module has failed.
LO
The P
sensor module has failed.
HI
The P
sensor module has failed.
LO
The P
sensor module has detected a
HI
pressure value that exceeds the configured
upper or lower alert value.
Communication between the P-Hi sensor
module and the electronics feature board
has been lost.
The P
Pressure reading has exceeded the
HI
sensor's maximum measurement range.
The P
sensor module has detected a
LO
pressure value that exceeds the configured
upper or lower alert value.
Communication between the P
module and the electronics feature board
has been lost.
The P
Pressure reading has exceeded the
LO
sensor's maximum measurement range.
The analog output of the ERS System is in
"fixed current mode" and is not
representative of the HART PV.
A sensor module is missing or not
detected.
Neither of the modules in the ERS System
are configured as the P
sensor.
HI
Recommended actions
Verify the process conditions and modify the
Analog Range Values if necessary.
Verify the measured DP is beyond the trip
limits. If necessary, modify the trip limits or
disable the diagnostic.
Replace the electronics feature board.
Verify the P
HI
the operating limits of the sensor. Replace
the P
sensor module if necessary.
HI
Verify the P
LO
the operating limits of the sensor. Replace
the P
sensor module if necessary.
LO
Verify the P
HI
the operating limits of the sensor. Replace
the P
sensor module if necessary.
HI
Verify the P
LO
the operating limits of the sensor. Replace
the P
sensor module if necessary.
LO
Verify the measured P
the trip limits. If necessary, modify the trip
limits or disable the diagnostic.
Verify the wiring between the P
and the electronics feature board and cycle
power to the entire ERS System. Replace the
P
module and/or electronics feature board
HI
if necessary.
Check the process for potential overpressure
conditions.
Verify the measured P
the trip limits. If necessary, modify the trip
limits or disable the diagnostic.
Verify the wiring between the P
sensor
and the electronics feature board and cycle
LO
power to the entire ERS System. Replace the
P
module and/or electronics feature board
LO
if necessary.
Check the process for potential overpressure
conditions.
Using a Field Communicator or AMS Device
Manager, disable "Loop Current Mode."
Verify that both sensors are connected and
properly wired.
Verify that both sensors are connected and
properly wired. Change the pressure
designation of one of the two modules to
"P
" using a Field Communicator or AMS
HI
Device Manager.
Reference Manual
00809-0100-4804, Rev CB
Module Temperature is within
Module Temperature is within
Module Temperature is within
Module Temperature is within
Pressure is beyond
HI
module
HI
Pressure is beyond
LO
module
LO
Troubleshooting

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 3051s

Table of Contents