Combined adjust-
ment procedure
step-by-step
Check & Adjust
The following table explains the most common settings. Refer to the stated chapter
for more information on screens.
Step
Description
1.
Turn on the instrument by pressing the
2.
Level up the instrument. Refer to "5.2 Instrument Setup" for more infor-
mation.
Make sure that THEO mode is active.
3.
Press
for approximately 5 seconds until SYSTEM INFO is active.
4.
Press CALIBR.
5.
Press NEW.
6.
± 5°
Aim with the telescope accurately at a
target at a distance of about 100 m.
The target must be positioned within
± 5° of the horizontal plane.
Builder
key.
169