Replacing the Seal Wash Pump

When
In case of wear of the seal wash pump
Parts required
p/n
5065-4445
5065-9978
Preparations
Remove the flow connections from and to the seal wash pump.
1
For removing the seal wash pump, press the clips (1) and
pull the pump to the front (2).
1290 Infinity Quaternary Pump User Manual
Description
Peristaltic pump with Pharmed tubing
Tubing, 1 mm i.d., 3 mm o.d., silicone, 5 m
2
Insert the pump clips to the holes in the module housing.
Maintenance

Replacing the Seal Wash Pump

9