Leak

Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and
board-mounted temperature-compensation sensor) are used by the leak
algorithm to determine whether a leak is present. When a leak occurs, the leak
sensor is cooled by the solvent. This changes the resistance of the leak sensor
which is sensed by the leak-sensor circuit on the main board.
Probable cause
1
2
3
4
1290 Infinity Quaternary Pump User Manual
Loose fittings.
Broken capillary.
Loose or leaking purge valve, inlet valve, or
outlet valve.
Defective pump seals.
Error Information
General Error Messages
Suggested actions
Ensure all fittings are tight.
Exchange defective capillaries.
Ensure pump components are seated correctly.
If there are still signs of a leak, exchange the
appropriate seal (purge valve, inlet valve, outlet
valve).
Exchange the pump seals.