Agilent Technologies 1290 Infinity User Manual page 168

Quaternary pump
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

9
Maintenance
Replacing the Multi-Channel Gradient Valve (MCGV)
7
Tighten the two screws (1, 2) and connect the cable to
its connector (3).
9
Reconnect the waste funnel with the waste tube holder
in the top cover. Insert waste tube in the holder in the
leak pan and clip tube to the MCGV cover.
168
8
Install the MCGV cover.
Next Steps:
10
Install the inlet weaver, see
Weaver"
on page 151.
11
Reconnect solvent tubes for channels A-D from the
MCGV to the degasser outlets.
1290 Infinity Quaternary Pump User Manual
"Replacing the Inlet