Leak And Waste Handling - Agilent Technologies 1290 Infinity User Manual

Quaternary pump
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

Leak and Waste Handling

Toxic, flammable and hazardous solvents, samples and reagents
WA R N I N G
The handling of solvents, samples and reagents can hold health and safety risks.
Recommendations for Solvent Cabinet
N O T E
For details, see the usage guideline for the Agilent 1200 Infinity Series Solvent Cabinets.
For details on correct installation, see section Installation Information on
Leak and Waste Handling in the service manual.
1290 Infinity Quaternary Pump User Manual
When working with these substances, observe appropriate safety procedures (for
example by wearing goggles, safety gloves and protective clothing) as described in
the material handling and safety data sheet supplied by the vendor, and follow
good laboratory practice.
The volume of substances should be reduced to the minimum required for the
analysis.
Do not operate the instrument in an explosive atmosphere.
Never exceed the maximal permissible volume of solvents (6 L) in the solvent
cabinet.
Do not use bottles that exceed the maximum permissible volume as specified in
the usage guideline for the Agilent 1200 Infinity Series Solvent Cabinets.
Arrange the bottles as specified in the usage guideline for the solvent cabinet.
A printed copy of the guideline has been shipped with the solvent cabinet,
electronic copies are available on the Internet.
Ground the waste container.
The residual free volume in the appropriate waste container must be large enough
to collect the waste liquid.
Check the filling level of the waste container regularly.
To achieve maximal safety, check regularly for correct installation.
Do not use solvents with an auto-ignition temperature below 200 °C (392 °F).
Using the Module

Leak and Waste Handling

4
55

Advertisement

Table of Contents
loading

Table of Contents