Error Information
Pump Error Messages

Timeout: Wait for run volume

Error ID: 29181
A target condition (run volume, which is the volume delivered since the
method run start) has been sent to the instrument which should have been
reached within an expected time frame but didn't. Either the limit, time frame
or the current value of the variable has been modified later directly or
indirectly (for example the flow rate).
Probable cause
1

Timeout: Wait for Volume

Error ID: 29182
A target condition (volume, which is the delivered flow since the limit has
been set) has been sent to the instrument which should have been reached
within an expected time frame but didn't. Either the limit, time frame or the
current value of the variable has been modified later directly or indirectly (for
example the flow rate).
Probable cause
1
128
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Suggested actions
Verify control software, macros, manual
commands.
Suggested actions
Verify control software, macros, manual
commands.
1290 Infinity Quaternary Pump User Manual