9
Maintenance
Installing the Jet Weaver
3
Insert the Jet Weaver to the opening in the front panel
(1) and push it down (2).
5
Connect the inlet capillary of the Jet Weaver to port 2 of
the Multi Purpose Valve. Connect the outlet capillary to
port 1.
162
4
Mount both capillary connections to the Jet Weaver
observing the correct orientation.
6
Configure the Jet Weaver as mixer in the user interface,
see
"Instrument Configuration"
1290 Infinity Quaternary Pump User Manual
on page 61.