Agilent 1290 Infinity
Thermostat
User Manual
Agilent Technologies

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 1290 Infinity

  • Page 1 Agilent 1290 Infinity Thermostat User Manual Agilent Technologies...
  • Page 2 1987) or any equivalent agency regulation or contract clause. Use, duplication or dis- closure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will 1290 Infinity Thermostat User Manual...
  • Page 3: Table Of Contents

    Transporting the Thermostatted Autosampler 4 Troubleshooting and Diagnostics Agilent Lab Advisor Software Overview of the Thermostat’s Indicators and Test Functions Status Indicators 5 Error Information What Are Error Messages General Error Messages Thermostat Error Messages 1290 Infinity Thermostat User Manual...
  • Page 4 Analog Cables Remote Cables BCD Cables Auxiliary Cable CAN/LAN Cables External Contact Cable RS-232 Cables 9 Appendix Safety Symbols The Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC) Radio Interference Sound Emission Agilent Technologies on Internet 1290 Infinity Thermostat User Manual...
  • Page 5: Introduction

    1290 Infinity Thermostat User Manual Introduction Introduction to the Thermostat Thermostat Operation Electrical Connections Serial Number Information (ALL) Rear View of the Module Early Maintenance Feetback Agilent Technologies...
  • Page 6: Introduction To The Thermostat

    Introduction to the Thermostat Introduction to the Thermostat The Agilent 1290 Infinity Thermostat is designed for use with other modules of the Agilent 1200 Infinity Series or with other LC systems if adequate remote control inputs and outputs are available. The thermostat is controlled from the Agilent 1200 Infinity Series Instant Pilot or from your Agilent control software for LC systems.
  • Page 7 Introduction Introduction to the Thermostat Figure 1 Overview of the Thermostat 1290 Infinity Thermostat User Manual...
  • Page 8: Thermostat Operation

    A fan draws air from the sample tray area and blows it through the channels of the heating/cooling module. The fan speed is determined according to the environmental conditions (e.g., ambient temperature, humidity). In the 1290 Infinity Thermostat User Manual...
  • Page 9 The fan speed is controlled according to the temperature of the Peltier elements. During cooling condensation will appear in the heating/cooling module. The condensed water will be guided out of the thermostat. 1290 Infinity Thermostat User Manual...
  • Page 10: Electrical Connections

    There are no externally accessible fuses, because automatic electronic fuses are implemented in the power supply. Never use cables other than the ones supplied by Agilent Technologies to ensure proper N O T E functionality and compliance with safety or EMC regulations.
  • Page 11 • DE = Germany • JP = Japan • CN = China year and week of last major manufacturing change, e.g. 820 could be week 20 of 1998 or 2008 SSSSS real serial number 1290 Infinity Thermostat User Manual...
  • Page 12 Introduction Electrical Connections Rear View of the Module Figure 3 Electrical Connections 1290 Infinity Thermostat User Manual...
  • Page 13: Early Maintenance Feetback

    EMF limits, and then reset the EMF counters to zero. The next time the EMF counters exceed the new EMF limits, the EMF flag will be displayed, providing a reminder that maintenance needs to be scheduled. 1290 Infinity Thermostat User Manual...
  • Page 14 Introduction Early Maintenance Feetback 1290 Infinity Thermostat User Manual...
  • Page 15: Site Requirements And Specifications

    1290 Infinity Thermostat User Manual Site Requirements and Specifications Site Requirements Power Consideration Power Cords Bench Space Environment Physical Specifications Performance Specifications Agilent Technologies...
  • Page 16: Site Requirements

    In case of emergency it must be possible to disconnect the instrument from the power line at any time. ➔ Make sure the power connector of the instrument can be easily reached and unplugged. ➔ Provide sufficient space behind the power socket of the instrument to unplug the cable. 1290 Infinity Thermostat User Manual...
  • Page 17: Power Cords

    Never operate your instrumentation from a power outlet that has no ground connection. ➔ Never use a power cord other than the Agilent Technologies power cord designed for your region. Use of unsupplied cables WA R N I N G Using cables not supplied by Agilent Technologies can lead to damage of the electronic components or personal injury.
  • Page 18: Bench Space

    Do not store, ship or use your module under conditions where temperature fluctuations could cause condensation within the module. ➔ If your module was shipped in cold weather, leave it in its box and allow it to warm slowly to room temperature to avoid condensation. 1290 Infinity Thermostat User Manual...
  • Page 19: Physical Specifications

    WA R N I N G Using the autosampler at high environmental temperatures may cause the rear panel to become hot. ➔ Do not use the autosampler at environmental temperatures higher than 50 °C (122 °F) 1290 Infinity Thermostat User Manual...
  • Page 20: Performance Specifications

    Site Requirements and Specifications Performance Specifications Performance Specifications Table 2 Performance specifications Agilent 1290 Infinity Thermostat Type Specification Temperature range: settable from 4 °C to 40 °C in 1 ° increments Temperature accuracy at ambient - 1 °C to + 4 °C at a setpoint of 4 °C temperatures <...
  • Page 21: Installing The G1330B Thermostat

    1290 Infinity Thermostat User Manual Installing the G1330B Thermostat Damaged Packaging Delivery Checklist Optimizing the Stack Configuration Installing the G1330B Thermostat Stage 1: Preparing the Thermostat and Autosampler Stage 2: Power Cable and Interface Cable Connection Stage 3: Flow Connections...
  • Page 22: Damaged Packaging

    The delivery checklist are shown in Table 3 on page 22. Please report missing or damaged parts to your local Agilent Technologies sales and service office. Table 3 G1330B Thermostat Checklist Description...
  • Page 23: Optimizing The Stack Configuration

    Figure 4 page 23 and Figure 5 on page 24. This configuration optimizes the system flow path, ensuring minimum delay volume. Figure 4 Recommended Stack Configuration (Front View) 1290 Infinity Thermostat User Manual...
  • Page 24 Installing the G1330B Thermostat Optimizing the Stack Configuration Figure 5 Recommended Stack Configuration (Rear View) 1290 Infinity Thermostat User Manual...
  • Page 25: Installing The G1330B Thermostat

    Agilent is required to evaluate if the instrument is in good condition or damaged. ➔ Notify your Agilent sales and service office about the damage. ➔ An Agilent service representative will inspect the instrument at your site and initiate appropriate actions. 1290 Infinity Thermostat User Manual...
  • Page 26: Stage 1: Preparing The Thermostat And Autosampler

    A suitable container must be provided and emptied regularly in order to avoid overflow. 1 Place the thermostat on the bench or in the stack. 2 Remove the front cover. Press the two snap fasteners on the sides of the cover and move it away. 1290 Infinity Thermostat User Manual...
  • Page 27 Make sure that the leak tube is fully fixed on the outlet. Figure 8 Condensation Leak outlet 1290 Infinity Thermostat User Manual...
  • Page 28: Stage 2: Power Cable And Interface Cable Connection

    This assures that the cold airstream from the thermostat is correctly guided to the tray area of the autosampler. 8 If there is no Agilent 1260/1290 Infinity Series module located beneath the thermostat connect the waste tube to the central waste exit of the autosampler and place in a waste vessel.
  • Page 29 Agilent 1200 Infinity Series Instant Pilot or ChemStation for LC for more information. In an Agilent 1290 Infinity system, the individual modules are connected by a CAN cable. N O T E The Agilent 1200 Infinity Series Instant Pilot can be connected to the CAN bus at any of the modules in the system.
  • Page 30 Installing the G1330B Thermostat Installing the G1330B Thermostat 7 Connect additional cables as required (see Figure 11 on page 31). Figure 10 Power Connectors and Safety Levers at Rear of thermostatted Autosampler. 1290 Infinity Thermostat User Manual...
  • Page 31 Installing the G1330B Thermostat Installing the G1330B Thermostat Figure 11 Cable Connections. 1290 Infinity Thermostat User Manual...
  • Page 32: Stage 3: Flow Connections

    1 Connect the pump outlet capillary to port 1 of the injection valve. 2 Connect column-compartment inlet capillary to port 6 of the injection valve. 3 Ensure that the waste tube is positioned inside the leak channel. Figure 12 Hydraulic Connections 1290 Infinity Thermostat User Manual...
  • Page 33: Stage 4: Installing The Sample Tray

    However when these trays are installed cooling or heating of the vials in the tray will not work. Half-trays can be installed in any combination enabling both 1.8 ml-and 6 ml-vials to be used simultaneously. 1290 Infinity Thermostat User Manual...
  • Page 34 The vial positions of the right-hand half tray begin at position 101 as follows: Left-hand 40-position tray: 1–40 Left-hand 15-position tray: 1–15 Right-hand 40-position tray: 101–140 Right-hand 15-position tray: 101–115 Figure 14 Numbering of Tray Positions. 1290 Infinity Thermostat User Manual...
  • Page 35: Stage 5: Installing Tray Cover And Front Cover

    The power switch stays depressed (1) and a green indicator lamp in the power switch is on N O T E when the autosampler is turned on. When the line power switch stands out (Ø) and the green light is off, the autosampler is turned off. 1290 Infinity Thermostat User Manual...
  • Page 36: Transporting The Thermostatted Autosampler

    Also, ensure the vial tray is secured in place with suitable packaging, otherwise the tray may become loose and damage internal components. 1290 Infinity Thermostat User Manual...
  • Page 37: Troubleshooting And Diagnostics

    1290 Infinity Thermostat User Manual Troubleshooting and Diagnostics Agilent Lab Advisor Software Overview of the Thermostat’s Indicators and Test Functions Status Indicators Power Supply Indicator Instrument Status Indicator Agilent Technologies...
  • Page 38: Agilent Lab Advisor Software

    Agilent Lab Advisor software may differ from the descriptions in this manual. For details refer to the Agilent Lab Advisor software help files. This manual provides lists with the names of Error Messages, Not Ready messages, and other common issues. 1290 Infinity Thermostat User Manual...
  • Page 39: Overview Of The Thermostat's Indicators And Test Functions

    In the event of an electronic, mechanical or hydraulic failure, the instrument generates an error message in the user interface. For details on error messages and error handling, please refer to the Agilent Lab Advisor software. 1290 Infinity Thermostat User Manual...
  • Page 40: Status Indicators

    Figure 16 Location of Status Indicators Power Supply Indicator The power supply indicator is integrated into the main power switch. When the indicator is illuminated (green) the power is ON. 1290 Infinity Thermostat User Manual...
  • Page 41: Instrument Status Indicator

    Usually, an error condition requires attention (for example, leak, defective internal components). An error condition always interrupts the analysis. 1290 Infinity Thermostat User Manual...
  • Page 42 Troubleshooting and Diagnostics Status Indicators 1290 Infinity Thermostat User Manual...
  • Page 43: Error Information

    1290 Infinity Thermostat User Manual Error Information What Are Error Messages General Error Messages Timeout Shut-Down Remote Timeout Synchronization Lost Leak Sensor Short Leak Sensor Open Compensation Sensor Open Compensation Sensor Short Fan Failed Leak Open Cover Cover Violation Thermostat Error Messages...
  • Page 44: What Are Error Messages

    (for example, repair, or exchange of consumables is necessary). In the event of such a failure, the red status indicator at the front of the module is switched on, and an entry is written into the module logbook. 1290 Infinity Thermostat User Manual...
  • Page 45: General Error Messages

    Check the logbook for the occurrence and A not-ready condition was present during a source of a not-ready condition. Restart the sequence or multiple-injection run for a analysis where required. period longer than the timeout threshold. 1290 Infinity Thermostat User Manual...
  • Page 46: Shut-Down

    Exchange the remote cable. Defective remote cable. Check the instrument for defects (refer to the Defective components in the instrument instrument’s documentation). showing the not-ready condition. 1290 Infinity Thermostat User Manual...
  • Page 47: Synchronization Lost

    If the current increases above the upper limit, the error message is generated. Probable cause Suggested actions Please contact your Agilent service Defective flow sensor. representative. Please contact your Agilent service Leak sensor incorrectly routed, being representative. pinched by a metal component. 1290 Infinity Thermostat User Manual...
  • Page 48: Leak Sensor Open

    If the resistance across the sensor increases above the upper limit, the error message is generated. Probable cause Suggested actions Please contact your Agilent service Defective main board. representative. 1290 Infinity Thermostat User Manual...
  • Page 49: Compensation Sensor Short

    Probable cause Suggested actions Please contact your Agilent service Fan cable disconnected. representative. Please contact your Agilent service Defective fan. representative. Please contact your Agilent service Defective main board. representative. 1290 Infinity Thermostat User Manual...
  • Page 50: Leak

    Broken capillary. Open Cover The top foam has been removed. Probable cause Suggested actions Please contact your Agilent service Foam not activating the sensor. representative. Please contact your Agilent service Dirty or defective sensor. representative. 1290 Infinity Thermostat User Manual...
  • Page 51: Cover Violation

    Probable cause Suggested actions The top foam was removed during operation. Please contact your Agilent service representative. Please contact your Agilent service Foam not activating the sensor. representative. 1290 Infinity Thermostat User Manual...
  • Page 52: Thermostat Error Messages

    Probable cause Suggested actions See Service Manual for Exchanging the One of the 4 heat-sink fans is defective. Heatsink Fans. See Service Manual for Exchanging Heat The cooling heating module fan is defective Exchanger Fan. 1290 Infinity Thermostat User Manual...
  • Page 53: Temperature Control Failed (1 - 4)

    Probable cause Suggested actions See Service Manual for Exchanging the Sensors. Temperature sensor is defective 1290 Infinity Thermostat User Manual...
  • Page 54: Lost Contact To Autosampler Thermostat

    Defective electronic board in autosampler or autosampler thermostat • Exchange TCA board in the thermostat • Contact the Agilent Technologies service organization Power Fail for Autosampler Thermostat Module The +36V that is generated in the autosampler thermostat power supply is checked by the autosampler electronics.
  • Page 55: Temperature Out Of Range

    Make sure that the ambient air temperature is Ambient temperature too high / low within its specified limits (4 °C to 55 °C) 1290 Infinity Thermostat User Manual...
  • Page 56: Bad Cooling / Heating Performance

    10 min for the autosampler thermostat peltier elements to equilibrate • Make sure that there is enough space on left and right side of the autosampler thermostat and that the air channel inlets and outlets are not blocked 1290 Infinity Thermostat User Manual...
  • Page 57: Maintenance

    1290 Infinity Thermostat User Manual Maintenance Warnings and Cautions Introduction to Maintenance Cleaning the Module Exchanging the Power Supply Fuses Agilent Technologies...
  • Page 58: Warnings And Cautions

    Never plug the power line back in when cover is removed. Sharp metal edges WA R N I N G Sharp-edged parts of the equipment may cause injuries. ➔ To prevent personal injury, be careful when getting in contact with sharp metal areas. 1290 Infinity Thermostat User Manual...
  • Page 59 C A U T I O N ➔ If you connect external equipment to the instrument, make sure that you only use accessory units tested and approved according to the safety standards appropriate for the type of external equipment. 1290 Infinity Thermostat User Manual...
  • Page 60: Introduction To Maintenance

    The module is designed for easy maintenance. Maintenance can be done from the front with module in place in the system stack. There are no serviceable parts inside. N O T E Do not open the module. 1290 Infinity Thermostat User Manual...
  • Page 61: Cleaning The Module

    WA R N I N G Liquid in the module electronics can cause shock hazard and damage the module. ➔ Do not use an excessively damp cloth during cleaning. ➔ Drain all solvent lines before opening any fittings. 1290 Infinity Thermostat User Manual...
  • Page 62: Exchanging The Power Supply Fuses

    2 Remove the power cable from the two modules. 3 Insert the flat head screwdriver in the fuse holder, slightly press and turn counter clockwise to release the fuse holder from the socket. 4 Pull the fuse holder out of the socket. 1290 Infinity Thermostat User Manual...
  • Page 63 5 Remove the fuse from the fuse holder. 6 Insert a new fuse in the fuse holder. 7 Reinsert the fuse holder and fix with the screwdriver. 8 Reinsert the power cables. 9 Switch ON the power switch. 1290 Infinity Thermostat User Manual...
  • Page 64 Maintenance Exchanging the Power Supply Fuses 1290 Infinity Thermostat User Manual...
  • Page 65: Parts For Maintenance

    1290 Infinity Thermostat User Manual Parts for Maintenance Main Assemblies (External Parts) Accessory Kit G1330-68755 HPLC System Tool Kit (G4203-68708) Foam Parts Plastic Parts Agilent Technologies...
  • Page 66: Main Assemblies (External Parts)

    Parts for Maintenance Main Assemblies (External Parts) Main Assemblies (External Parts) Figure 17 Main Assemblies 1290 Infinity Thermostat User Manual...
  • Page 67 Description 2110-0015 Fuses T2.5 A/250V (CSA, UL listed) 2110-0029 Fuse TCA - Board (T3 A/250 V; CSA, UL listed) 5065-9982 Plastics kit (includes base, top, left and right sides) G1330-81600 Cable, autosampler to ALS thermostat 1290 Infinity Thermostat User Manual...
  • Page 68: Accessory Kit G1330-68755

    Parts for Maintenance Accessory Kit G1330-68755 Accessory Kit G1330-68755 Description G1330-67300 Waste Tube Assy CA G1330-81600 Cable, autosampler to ALS thermostat G1330-90120 hints for use Autosampler Thermostat 1290 Infinity Thermostat User Manual...
  • Page 69: Hplc System Tool Kit (G4203-68708)

    Parts for Maintenance HPLC System Tool Kit (G4203-68708) HPLC System Tool Kit (G4203-68708) HPLC System Tool Kit (p/n G4203-68708) contains some accessories and tools needed for installation and repair of the module. 1290 Infinity Thermostat User Manual...
  • Page 70: Foam Parts

    Parts for Maintenance Foam Parts Foam Parts Figure 18 Foam Parts Table 4 Main Assemblies Item Description Part Number Top Foam p/n G1330-40102 Bottom Foam p/n G1330-40103 1290 Infinity Thermostat User Manual...
  • Page 71: Plastic Parts

    Parts for Maintenance Plastic Parts Plastic Parts Figure 19 Plastic Parts Item p/n Description G1330-68723 Cover kit (includes base, top, left and right) Frontcover Kit 5067-4691 5042-8567 Leak pan 1290 Infinity Thermostat User Manual...
  • Page 72 Parts for Maintenance Plastic Parts 1290 Infinity Thermostat User Manual...
  • Page 73: Cable Overview

    1290 Infinity Thermostat User Manual Cable Overview Cable Overview Analog Cables Remote Cables BCD Cables Auxiliary Cable CAN/LAN Cables External Contact Cable RS-232 Cables Agilent Technologies...
  • Page 74: Cable Overview

    Cable Overview Cable Overview Cable Overview Never use cables other than the ones supplied by Agilent Technologies to ensure proper N O T E functionality and compliance with safety or EMC regulations. Analog cables Description 35900-60750 Agilent module to 3394/6 integrators...
  • Page 75 It's also called "Null Modem Cable" with full handshaking where the wiring is made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9. 5181-1561 RS-232 cable, 8 m 1290 Infinity Thermostat User Manual...
  • Page 76: Analog Cables

    The other end depends on the instrument to which connection is being made. Agilent Module to 3394/6 Integrators p/n 35900-60750 Pin 3394/6 Pin Agilent Signal Name module Not connected Shield Analog - Center Analog + 1290 Infinity Thermostat User Manual...
  • Page 77 Pin BNC Pin Agilent Signal Name module Shield Shield Analog - Center Center Analog + Agilent Module to General Purpose p/n 01046-60105 Pin 3394/6 Pin Agilent Signal Name module Not connected Black Analog - Analog + 1290 Infinity Thermostat User Manual...
  • Page 78: Remote Cables

    Cable Overview Remote Cables Remote Cables One end of these cables provides a Agilent Technologies APG (Analytical Products Group) remote connector to be connected to Agilent modules. The other end depends on the instrument to be connected to. Agilent Module to 3396A Integrators...
  • Page 79 5 - Pink Not connected 6 - Yellow 6 - Yellow Power on High 7 - Red 7 - Red Ready High 8 - Green 8 - Green Stop 9 - Black 9 - Black Start request 1290 Infinity Thermostat User Manual...
  • Page 80 Prepare run 3 - Gray Start 4 - Blue Shut down 5 - Pink Not connected 6 - Yellow Power on High 7 - Red Ready High 8 - Green Stop 9 - Black Start request 1290 Infinity Thermostat User Manual...
  • Page 81: Bcd Cables

    BCD 4 Black BCD 0 Orange BCD 3 BCD 2 Brown BCD 1 Gray Digital ground Gray Gray/pink BCD 11 Red/blue BCD 10 White/green BCD 9 Brown/green BCD 8 not connected not connected + 5 V 1290 Infinity Thermostat User Manual...
  • Page 82 Agilent Module to 3396 Integrators p/n 03396-60560 Pin 3396 Pin Agilent Signal Name BCD Digit module BCD 5 BCD 7 BCD 6 BCD 4 BCD0 BCD 3 BCD 2 BCD 1 Digital ground + 5 V 1290 Infinity Thermostat User Manual...
  • Page 83: Auxiliary Cable

    Agilent vacuum degasser. The other end is for general purpose. Agilent Vacuum Degasser to general purposes p/n G1322-81600 Color Pin Agilent Signal Name 1100 White Ground Brown Pressure signal Green Yellow Grey DC + 5 V IN Pink Vent 1290 Infinity Thermostat User Manual...
  • Page 84: Can/Lan Cables

    CAN cable, Agilent module to module, 1 m LAN Cables Description 5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection) 5023-0202 Twisted pair network cable, shielded, 7 m (for point to point connection) 1290 Infinity Thermostat User Manual...
  • Page 85: External Contact Cable

    Yellow EXT 2 Grey EXT 3 Pink EXT 3 Blue EXT 4 EXT 4 Black Not connected Violet Not connected Grey/pink Not connected Red/blue Not connected White/green Not connected Brown/green Not connected White/yellow Not connected 1290 Infinity Thermostat User Manual...
  • Page 86: Rs-232 Cables

    It's also called "Null Modem Cable" with full handshaking where the wiring is made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9. 5181-1561 RS-232 cable, 8 m 1290 Infinity Thermostat User Manual...
  • Page 87 1290 Infinity Thermostat User Manual Appendix Safety Symbols The Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC) Radio Interference Sound Emission Agilent Technologies on Internet Agilent Technologies...
  • Page 88: Appendix Safety Symbols

    C A U T I O N alerts you to situations that could cause loss of data, or damage of equipment. ➔ Do not proceed beyond a caution until you have fully understood and met the indicated conditions. 1290 Infinity Thermostat User Manual...
  • Page 89 Make sure that only fuses with the required rated current and of the specified type (normal blow, time delay, and so on) are used for replacement. The use of repaired fuses and the short-circuiting of fuse holders must be avoided. 1290 Infinity Thermostat User Manual...
  • Page 90 When working with solvents please observe appropriate safety procedures (e.g. goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet by the solvent vendor, especially when toxic or hazardous solvents are used. 1290 Infinity Thermostat User Manual...
  • Page 91: The Waste Electrical And Electronic Equipment (Weee) Directive (2002/96/Ec)

    With reference to the equipment types in the WEEE Directive Annex I, this product is classed as a Monitoring and Control Instrumentation product. Do not dispose off in domestic household waste N O T E To return unwanted products, contact your local Agilent office, or see www.agilent.com more information. 1290 Infinity Thermostat User Manual...
  • Page 92: Radio Interference

    Radio Interference Radio Interference Cables supplied by Agilent Technologies are screened to provide optimized protection against radio interference. All cables are in compliance with safety or EMC regulations. Test and Measurement If test and measurement equipment is operated with unscreened cables, or used for measurements on open set-ups, the user has to assure that under operating conditions the radio interference limits are still met within the premises.
  • Page 93: Sound Emission

    This product has a sound pressure emission (at the operator position) < 70 dB. • Sound Pressure Lp < 70 dB (A) • At Operator Position • Normal Operation • According to ISO 7779:1988/EN 27779/1991 (Type Test) 1290 Infinity Thermostat User Manual...
  • Page 94: Agilent Technologies On Internet

    Agilent Technologies on Internet For the latest information on products and services visit our worldwide web site on the Internet at: http://www.agilent.com Select Products/Chemical Analysis It will provide also the latest firmware of the modules for download. 1290 Infinity Thermostat User Manual...
  • Page 95 Diagnostic software cable dimensions failure bench space fan failed flow path electrical connections frequency range cable descriptions of fuses analog electronic waste auxiliary early maintenance feedback general error messages environment external contact 1290 Infinity Thermostat User Manual...
  • Page 96 RS-232C feedback cable vial tray message vial racks cover violation vials ignition without cover safety class I 1290 Infinity Thermostat User Manual...
  • Page 97 Index voltage range warnings and cautions waste electrical and electronic equipment WEEE directive weight 1290 Infinity Thermostat User Manual...
  • Page 98 In This Book This manual contains information about the Agilent 1290 Infinity Thermostat. This manual describes the following: • introduction, • site requirements and specifications, • installing the thermostatted autosampler, • optimizing performance, • troubleshooting and diagnostics, • maintenance, •...

Table of Contents