Daily / Weekly tasks

Daily tasks
• Replace solvents and solvent bottles for mobile phases based on
• Replace solvents and solvent bottles for organic mobile phase latest every
• Check presence of seal wash solvent.
• Purge each channel with fresh solvent at 2.5 – 3 mL/min for 5 min.
• Equilibrate your system with composition of your application for 15 min.
Weekly tasks
• Change seal wash solvent (10 % / 90 % isopropanol/water) and bottle.
• Flush all channels with water at 2.5 – 3 mL/min for 5 min to remove salt
• Inspect solvent filters for dirt or blockages. Clean or exchange if no flow is
56
water/buffer.
second day.
Use conditioning for 1290 Infinity II Pumps and G7104C.
deposits if buffer applications were used.
coming out of the solvent line when removed from the degasser inlet.