Oxi 3315
Configuring the
automatic storage
function
Settings
Starting the auto-
matic storage func-
tion
ba75944e02
03/2012
7.2
Automatic storage at intervals
The storage interval (Interval) determines the time interval between automatic
storage processes. Each storage process transmits the current dataset to the
interface at the same time.
1.
Press the <STO_> key.
The menu for automatic storage appears.
Automatic data storage
ID number
Interval
Duration
continue
0d03h00min
0
Back
1 Adjusted entire storage duration
2 Max. available storage duration
3 Graphical display of the storage usage
You can configure the automatic storage function with the following settings:
Menu item
Possible setting
ID number
1 ... 10000
Interval
1 s, 5 s, 10 s,
30 s, 1 min,
5 min, 10 min,
15 min, 30 min,
60 min
Duration
1 min ... x min
To start the automatic data storage function, select continue with <><> and
1
30 s
180 min
1d17h33min
01.09.2011
08:00
Description
ID number for the dataset series.
Storage interval.
The lower limit of the storage interval
can be restricted by the number of free
storage locations.
The upper limit is restricted by the stor-
age duration.
Storage duration.
Specifies after which time the auto-
matic storage should be terminated.
The lower limit of the storage duration
is restricted by the storage interval.
The upper limit is restricted by the
number of free storage locations.
Data storage
1
2
3
39