wtw Oxi 3315 Operating Manual page 32

Digital meter for digital ids dissolved oxygen sensors
Hide thumbs Also See for Oxi 3315:
Table of Contents

Advertisement

Dissolved oxygen
®
FDO
925 calibra-
tion evaluation
Calibration record
(USB output)
32
calibration. The evaluation appears on the display and in the calibration record.
Display
Error
Error elimination (see section 10 W
page 48)
CALIBRATION Ox
Calibration date01.09.2011 07:43:33
FDO 925
Ser. no. 10146858
SC-FDO 925
Relative slope
Sensor
Calibration record
+++
++
+
Error
HAT TO DO IF
10158765
0.98
+++
Oxi 3315
Relative slope
S = 0.94 ... 1.06
S = 0.92 ... 0.94
or
S = 1.06 ... 1.08
S = 0.90 ... 0.92
or
S = 1.08 ... 1.10
S < 0.90
or
...,
S > 1.10
ba75944e02
03/2012

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents