wtw Oxi 3315 Operating Manual page 30

Digital meter for digital ids dissolved oxygen sensors
Hide thumbs Also See for Oxi 3315:
Table of Contents

Advertisement

Dissolved oxygen
30
Before calibrating via a comparison measurement, the sensor
should be calibrated in the air calibration vessel.
Proceed as follows to calibrate the instrument:
1.
Connect the D.O. sensor to the measuring instrument.
2.
Immerse the D.O. sensor in the reference solution.
3.
Start the calibration with <CAL>.
O
2
FDO Comp
4.
Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
5.
Wait for the end of the AutoRead measurement or accept the calibration
value with <ENTER>.
The factor that was set last is displayed.
O
2
FDO Comp
7.92
mg/l
25.1
°C
22.09.2008
08:00
7.92
mg/l
0.973
22.09.2008
08:00
Oxi 3315
ba75944e02
03/2012

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents