Firmware Update; Firmware Update For The Meter Oxi 3315 - wtw Oxi 3315 Operating Manual

Digital meter for digital ids dissolved oxygen sensors
Hide thumbs Also See for Oxi 3315:
Table of Contents

Advertisement

Firmware update

52
12
Firmware update

12.1 Firmware update for the meter Oxi 3315

You can find availble firmware update files for your meter on the internet. With
the "Firmware Update " program and a PC you can update the firmware of the
Oxi 3315 to the newest version.
For the update you have to connect the meter to a PC.
For the update via the USB-B interface, the following is required:
 a free USB interface (virtual COM port) on the PC
 the driver for the USB interface (on the enclosed CD-ROM)
 the USB cable (included in the scope of delivery of the Oxi 3315).
1.
Install the downloaded firmware update on a PC.
An update folder is created in the Windows start menu.
If an update folder already exists for the meter (or meter type), the new
data are displayed there.
2.
In the windows start menu, open the update folder and start the firmware
update program for the meter
3.
Using the USB interface cable, connect the Oxi 3315 to a USB interface
(virtual COM port) of the PC.
4.
Switch on the Oxi 3315.
5.
In the firmware update program, start the update process with OK.
6.
Follow the instructions of the firmware update program.
During the programming process, a corresponding message and a prog-
ress bar (in %) are displayed.
The programming process takes up to 5 minutes. A terminatory mes-
sage is displayed after a successful programming process. The firmware
update is completed.
7.
Disconnect the Oxi 3315 from the PC.
The Oxi 3315 is ready for operation again.
After switching the meter off and on you can check whether the meter has taken
over the new software version (see Y
VERSION IS IN THE METER OR
OU WANT TO KNOW WHICH SOFTWARE
IDS
,
49).
SENSOR
PAGE
Oxi 3315
ba75944e02
03/2012