Self-Test Error When The Power Is Turned On - YOKOGAWA MT210 User Manual

Digital manometer
Table of Contents

Advertisement

No. Type of Error
33
Store error
38
Store error
50
Error in measured
pressure value
51
Error in measured
pressure value
52
Error in measured
pressure value
53
DMM measurement error DMM A/D conversion timeout.
54
DMM measurement error DMM measurement update error.
55
DMM measurement error DMM measurement overrange.
56
DMM measurement error %ERROR calculation overflow.

Self-Test Error When the Power is Turned ON.

No. Type of Error
60 Setting information
backup error (excluding
communications settings)
61 Communications settings
information backup error
62 Measurement data backup
error
63 Zero calibration backup value
failure (absolute pressure
models only)
83 EEPROM error (DMM
calibration)
84 EEPROM error (D/A output
calibration value)
90 Hardware error
91 Pressure measurement
section calibration data
failure (sensor section)
92 Calibration error in the
pressure measurement
(calculation) section
93 Setting value error in the
pressure measurement
section (sensor section)
94 Setting value error in the
pressure measurement
(calculation) section
• If an error 60–63, 83, or 84 occur when turning ON the power, the error will be displayed until a certain key is pressed.
IM 767361-01E
Cause
Internal memory full
Attempted to store measurement data
zunder an already existing data number.
Measured pressure value update error.
"---------" appears in the upper part of the
screen.
Overrange occurred in the measured
pressure value. "--oL--" is displayed in the
upper part of the screen.
Overflow occurred in the measured
pressure value.
"--oF--" is displayed in the upper part of
the screen (or [SPn.Err] when in % display
mode).
"---------" appears in the lower part of the
screen,
"--oL--" appears in the lower part of the
screen.
"--oF--" is displayed in the lower part of the
screen.
Cause
Panel setting information other than
communications settings were corrupted.
File is damaged.
Measurement data saved to the internal Initialize
the data.
memory was corrupted.
Values for Zero CAL were corrupted.
(767357/767367/767387)
EEPROM values were corrupted.
EEPROM values were corrupted.
Hardware failure in the pressure measurement
section.
Calibration data from the pressure sensor were
corrupted.
Calibration data from the pressure calculation
section were corrupted.
Setting data from the pressure sensor were
corrupted.
Failure in the setting values of the pressure
calculation section.
Troubleshooting, Maintenance, and Inspection
Corrective Action
Delete unneeded measurement data.
Delete the existing data, or use an
unused data number.
Service required if this continues to be
displayed.
Input a pressure within 120% of the
measurement range.
If [SPn.Err] is displayed, reenter the
correct pressure value for 0%, 100%.
Service required.
Service required if this continues to be
displayed. (Service is required if, when
using the OE communications command
to readout the error message, that
message to be displayed.)
Input a pressure value within the
displayed range.
Check whether the pressure percentage,
voltage, and current values are all
correctly displayed.
Corrective Action
Initialize the settings.
Initialize the settings.
Initialize the Zero CAL values. 23
DMM calibration is required.
Service is required.
Service required.
Service required.
Service required.
Service required.
Service required.