Data Output Format; Header Section - YOKOGAWA MT210 User Manual

Digital manometer
Table of Contents

Advertisement

RS-232 (Response to the <ESC>S Command)

Data Output Format

Output Format for Measurement and Computation Data (Response to the "OD" Command)
Data Format

Header Section

IM 767361-01E
DI08
DI07
DI06
0
1
ERROR
(fixed)
(fixed)
For DIO1 to DIO6, the status byte format is identical to that used for the GP-IB
interface. For details, refer to page App-12.
Note
After the status byte responds to the serial poll from the controller, this bit is reset to 0. Until
the serial poll is responded to, each interrupt is stored, and if more interrupts occur thereafter,
they are added.
• To prevent the status byte being affected by a computation END, syntax ERROR, or OVER,
use the "IM" command to set a value with the corresponding bits set to 0.
Each data consists of a 4-byte header and 8 bytes of data.
You can specify whether a header is output using the H command.
Header
Data
The header section consists of four bytes (h1 to h4).
h1 h2 h3 h4
h1 to h3:
KPA:
Pressure (kPa)
KGF:
Pressure (kgf/cm
MHO:
Pressure (mmH
0)
2
MHG:
Pressure (mmHg)
IHO:
Pressure (inH
0)
2
IHG:
Pressure (inHg)
PSI:
Pressure (psi)
PP
: Pressure %
DMM Header (MT220 Only)
h1 to h3: Data type
V
: DC voltage (V)
A
: DC current (mA)
PV
: DC voltage %
PA
: DC current %
ERR: Error %
DI05
DI04
DI03
24 VOUT
Syntax
OVER
-OL-
Error
Data type h4:
N: Normal measured data
2
)
I:
Overrange data
O: Computation overflow
E: No data
h4: Data state
N:
Normal measured data
I:
Overrange data
O: Computation overflow
F: No DMM data when DMM function is off
E: No data
Appendix
DI02
DI01
Auto
Calcula-
store
tion
END
END
indicates a space.
Data state
App-13