DMM Function

% Display

Setting 0% and 100% by Entering the Desired Numerical Value (Manual Setting)
IM 767361-01E
Adjusting and Calibrating the Differential Pressure Transmitter using the MT220
1.
If nothing is displayed in the lower part of the screen, press the ON/OFF under the
DMM group to turn the DMM function ON.
2.
Press DC5 V/20 mA to set the mark to either V or mA.
3.
To cancel DMM measurement, press the ON/OFF under the DMM group again to
turn the DMM function OFF.
Note
The display range for the measured values are shown below.
Voltage –6.0000 V to +6.0000 V
Current –24.000 mA to +24.000 mA
If 0% and 100% pressure values have been set, the measured values can be
displayed as a percentage (%) relative to these 0% and 100% values. 0% and
100% values can be set by directly entering a desired numerical value or by
applying actual pressure.
If % display mode is active during measurement of current or voltage using the
DMM function, the measured value is displayed as a percentage: 1 to 5 V or 4
to 20 mA being equivalent to 0 to 100%. The display range for pressure and
voltage/current values is shown below.
Pressure: 999.99%.
value exceeds the specified number of display digits.
Current/voltage: 0 to 24 mA/0 to 6 V (–25 to 125%).
lower part of the screen if the input exceeds this range.
Example
When 20 kPa is set as 0% and 100 kPa as 100%:
If the input pressure is 40 kPa, 25% will be displayed in the upper part of the
screen. 25% (8 mA) will also be displayed in the lower part of the screen if the
output current from the differential pressure transmitter is correct..
100%
25
0
20
40
100 kPa
1.
Press MENU.
Use the
and
to select
pressure setting screen.
2.
Use the
,
and
to enter the value corresponding to 0%, then press ENTER/
LOCAL.
3.
To set the value corresponding to 100% value, use the ,
ENTER/LOCAL.
4.
Press MENU to return to the normal measurement screen.
5.
Press the %ERROR to activate % display mode.
is displayed in the upper part of the screen if the
100%
37.5
25
0
4
8 10
20 mA
, then press ENTER/LOCAL to activate the 0%
is displayed at the
and
then press
41