Input Function

IM 767361-01E
When working with absolute pressure models, zero calibration is very complex, so use the
relative function to correct errors if the operating position differs from that in which zero
calibration was performed.
In the case of the absolute pressure model, zero calibration values cannot be changed by
initializing setup information using the RC communications command or the ENTER/LOCAL
after the power switch is turned ON.
In the case of the absolute pressure model, zero calibration values are not initialized even if
error code 60 (set-up information back-up error) occurs. However, they are initialized after a
voltage drop in the back-up battery. In this case, error 63 will be displayed.
• Never apply a pressure exceeding the permitted input pressure to the
input port, otherwise damage to the instrument may result.
The allowable input pressure range is shown below.
Model
Allowable Input Pressure Range
767351
2.7 kPa abs to 500 kPa gauge
767353
2.7 kPa abs to 500 kPa gauge
767355
2.7 kPa abs to 3000 kPa gauge
767356
2.7 kPa abs to 4500 kPa gauge
767357
1 Pa abs to 500 kPa abs
767361
2.7 kPa abs to 500 kPa gauge
767363
2.7 kPa abs to 500 kPa gauge
767365
2.7 kPa abs to 3000 kPa gauge
767366
2.7 kPa abs to 4500 kPa gauge
767367
1 Pa abs to 500 kPa abs
767370
1 Pa abs to
767371
2.7 kPa abs to 500 kPa gauge
767372
2.7 kPa abs to 500 kPa gauge
767373
2.7 kPa abs to 1000 kPa gauge
767381
2.7 kPa abs to
767383
2.7 kPa abs to 500 kPa gauge
767385
2.7 kPa abs to 3000 kPa gauge
767386
2.7 kPa abs to 4500 kPa gauge
767387
1 Pa abs to 500 kPa abs
*1: The difference between L and H is less than or equal to 50 kPa.
• Make sure the following input pressure is not exceeded when using the
767370 or 767381. If exceeded, damage to the the sensor section may
result.
767370: A difference in pressure between the High and Lo sides of less
than 50 kPa.
767381: A difference to the atmospheric pressure of less than 50 kPa.
CAUTION
Display
–12.0000 to 12.0000 kPa
Up to 156.000 kPa
Up to 840.00 kPa
Up to 3600.00 kPa
Up to 156.000 kPa abs
–12.0000 to 12.0000 kPa
156.000 kPa or lower
840.00 kPa or lower
3600.00 kPa or lower
156.000 kPa abs or lower
*1
50 kPa gauge
–1.20000 to 1.20000 kPa
–12.0000 to 12.0000 kPa
–156.000 to 156.000 kPa
–156.00 to 840.00 kPa
50 kPa gauge
–12.0000 to 12.0000 kPa
156.000 kPa or lower
840.00 kPa or lower
3600.00 kPa or lower
156.000 kPa abs or lower
Pressure Measurement